Instruction Manual
DM08
6.2
Electrode and magnetic current line
For the separate design, the electrodes and magnetic current cable must be connected in addition.
CAUTION! Electric current!
Do not connect or disconnect the field coil cable before the primary power of the meter has been
disconnected!
Connect through the terminal strips in the junction boxes of the display electronics and the sensor.
Pin assignment:
Terminal Wire colour
Function
Magnetic current line (3 wire)
5
brown
Magnetic field current 1
6
white
Magnetic field current 2
7
green-yellow Potential equalization / PE
Electrode line (5 wire*)
22
red
Measuring ground
23
brown
Electrode 1
24
white
Electrode 2
* The wires green-yellow and blue are not connected.
Only use cables as specified in chapter "Connecting cables“.
The outer shield is grounded by means of special EMC-compliant cable glands at both ends of
the cable.
The inner shields are connected to terminal 7 and 22 respectively.
- 20 -