Basically, the measuring principle does not depend on the flow profile. You should use
straight inlet and outlet sections according to the nominal width (DN). An inflow path of 5 x
DN and an outflow zone of at least 2 to 3 x DN is required.
The flow sensor is only suitable for application in completely filled pipe systems. With a free
pipe outlet, the sensor should not be installed in pipe sections that could run empty (e.g.,
downpipes).
If the sensor must be installed in a down pipe, ensure that portion of the pipeline is always
filled 100% with the media.
Due to possible accumulation of gases, the sensor should not be installed at the highest
point of a pipeline.
It must be ensured, that the axes of the electrodes are running horizontally to avoid erro-
neous measurements due to deposits or air bubbles on the electrodes.
Also, note that in case of deviations from horizontal, the functioning of the Empty pipe de-
tection (p. 9) is limited.
- 14 -
PKP Prozessmesstechnik • 09/2019