102
-221 Settings conflict; FM deviation cannot exceed carrier
The frequency deviation cannot exceed the carrier frequency, or the
waveform generator will adjust the deviation to a maximum allowed
value.
-221 Settings conflict; FM deviation exceeds max frequency
The sum of the carrier frequency and the deviation cannot exceed the
maximum frequency for the chosen function plus 100 kHz (20.1 MHz
for sine and square, 300 kHz for ramp, and 5.1 MHz for arbitrary
waveforms). If you set the carrier to a value that is not valid, the
waveform generator will adjust the deviation to the maximum value
allowed.
-221 Settings conflict; PWM deviation decreased due to pulse
parameters
The PWM deviation must follow the limitation concerning the pulse
width or duty cycle, edge time, and period (see Ch 4.11 for these
limitations), or the waveform generator will adjust it automatically.
-221 Settings conflict; frequency forced duty cycle change
In the square wave function, if you change the frequency to a value
that cannot produce the current duty cycle, the duty cycle will be
automatically adjusted to the maximum value for the new frequency.
-221 Settings conflict; marker forced into sweep span
The marker frequency must be a value between the start and the stop
frequency.
-221 Settings conflict; selected arb is missing, changing
selection to default
If you delete an arbitrary waveform after storing its instrument state,
you will lose the waveform data and not able to output the waveform.
The built-in exponential-rise wave is outputted instead.
-221 Settings conflict; offset changed due to amplitude
When the existing offset voltage does not conform the following
Summary of Contents for G5100A