8 Troubleshooting
L-402 Mniature Stage
MP164E
Version: 1.2.0
23
Problem
Possible causes
Solution
Reduced positioning accuracy Warped base body
Mount the L-402 onto an even surface. The
recommended flatness of the surface is
≤ 10 µm.
When the L-402 is
mounted vertically: The
load exceeds the self-
locking of the drive.
Reduce the load. Make sure that the self-
locking (see specification of the holding
force in the data table) of the drive is not
exceeded (p. 27)).
Increased wear due to
small motions over a
long period of time
Perform a maintenance run over the entire
travel range (p. 21).
Functional impairment after
system modification
Controller was
replaced.
The LS402 was
replaced by another
model.
Controller from PI:
Load the parameters from the stage
database that correspond to the
combination of controller and the L-310
model.
Controller from a third-party supplier:
Check the operating parameters.
8
Troubleshooting