7000545 Revision A
Defibrillator
Waveform
Biphasic Truncated Exponential with voltage and duration
compensation for patient impedance.
Patient Impedance
Range
10 to 300 ohms. If the detected impedance is outside this range, the
user is prompted to check the electrode pads and connector. The
device will not deliver a shock if impedance is outside this range.
Output Energy
Configurable energy levels for Shock 1, Shock 2, and Shock 3 or
greater.
Adult mode: 150, 200, 300, or 360 joules.
Child mode: 35, 50, 75, or 90 joules.
Output Energy
Accuracy
10% of the energy setting into 50 ohms.
15% of the rated energy output into 25–175 ohms.
Rated Energy Output
Rated energy output is the nominal delivered energy based on the energy setting and patient impedance, as defined in
the following chart.
Waveform Shape and Measured Parameters