Title
Service Manual, LPCR2
Page
75 of
104
Doc
Type
Service Product Drawing
Doc
ID
3323299
Doc
Rev
B
Doc
State
Released
State Date
11/14/2017
7000545 Revision A
NOTE
:
1. AED Event Viewer will keep the AED events for 7 days.
2. This Data flow method requires LIFENET implementation.
Using DT EXPRESS Data Transfer Software to download an AED Event
Case 3 (non-connected CR2 or USB-only CR2):
For non-connected LIFEPAK CR2 devices, including USB-only, if customers purchase AED Post-Event Service, an
FSR will be sent on-site to use DT EXPRESS to manually download the AED Event from the
CR2
. The DT EXPRESS
version which supports the
CR2