Service Manual, LPCR2
Page
94 of
104
Doc
Type
Service Product Drawing
Doc
ID
3323299
Doc
Rev
B
Doc
State
Released
State Date
11/14/2017
7000545 Revision A
LIFENET Connection Error
An error has occurred while attempting to connect
to LIFELINK or LIFENET System. Contact
SmartDesk for support.
Connection Established
Once LIFENET connection has been established.
First HTTP response is an OK. Start LIFENET
Check-In after the completion of "Connection
Established"
This May Take Several
Minutes
After the completion of "Connection Established"
and before performing LIFENET Check-In. During
LIFENET Check-In, device will beep every 5
seconds.
Update In Progress
When software update is initiated with LIFNET
Rebooting
Once software update is complete and ready for
reboot. Software shall reboot after this prompt.
When reboot is done, "Mobile Connection Initiated"
prompt plays again.
Communication Complete
After LIFENET Check-In completes.
Connection Lost
If current LIFENET Check-In session aborts after a
successful LIFENET connection has been
established. Check cellular signal from a smart
phone and try again.
Powering Off
Upon termination of Cellular Troubleshooting.
There are additional voice prompts that are used during a cardiac arrest event. See Operator Instruction for the
complete list of voice prompts.
Error Codes
Overview
: If any of following error codes occurs on the device, the device status will change to “Not Ready”. Follow
recommended corrective action code to handle the errors.
Corrective Action Code Key
Code
Description
1
Replace Unit
Error Code
Hex Id
Event Name
Service
Corrective
Action
0x40012
AOKLOGGEREVENT_HighChrgRate
1
0x40013
AOKLOGGEREVENT_LowChrgRate
1
0x40014
AOKLOGGEREVENT_OverVolt
1
0x40017
AOKLOGGEREVENT_MajorDiff
1
0x40020
AOKLOGGEREVENT_AokReset
1
0x40021
AOKLOGGEREVENT_SysReset
1