![Physio Control 805355 Cleaning And Sterilization Manual Download Page 5](http://html1.mh-extra.com/html/physio-control/805355/805355_cleaning-and-sterilization-manual_1553923005.webp)
Internal Defibrillation Handles and Electrodes
01/2010
HANDLING, CLEANING, AND STERILIZATION GUIDELINES
CAUTION:
• Use one sterilization method for the life of the product. Using different sterilization methods on the same accessory may damage
the accessory and limit the effectiveness of the additional sterilization cycles.
• The useful life of internal defibrillation handles and electrodes is affected by the number of sterilization cycles rather than by the
age of the handles and electrodes.
• Take great care to protect each electrode during transport, use, cleaning, and sterilization, to prevent damage to electrode
surfaces. Prevent electrodes from bumping one another, or other equipment. Handle electrodes individually, not as a group.
Wrap each electrode individually for sterilization.
The following paragraphs provide current handling, cleaning, and sterilization guidelines for internal defibrillation handles and
electrodes (paddles).
Please keep this guideline document with your LIFEPAK defibrillator operating instructions manual and in
your sterilization department.
The accessories covered by this guideline are:
• Internal handles with discharge control (PN 3010901 and PN 805249)
• Internal electrodes (PN 805355)
During Surgical Set-up:
• Examine handles, cables, and connectors for damage or signs of wear (such as loose cable connections, damaged pins,
exposed wires, or cable connector corrosion).
• Before connecting the handles to a defibrillator, press the shock button on the handle to check for actuation. An audible click
should be heard from the shock button.
• Examine the electrodes for scratched, pitted, or chipped surfaces, and for bubbled, scratched, or chipped epoxy coating.
If any of these conditions are found, or if an audible click is not heard from the shock button during the actuation check, remove the
affected component from use immediately.
After Each Use and Prior to Sterilizing:
• Remove electrode(s) from handle(s).
• Manually wipe or rinse electrodes, handles, cables, and connectors with Enzol
®
or chemically equivalent mild enzymatic
detergent using a damp sponge or towel. A brush may be used on the handle, including its cable and connector.
Do not use a brush on the electrodes. Do not immerse or soak (except for the removable internal electrodes).
• Rinse all parts thoroughly with deionized water.
• Dry thoroughly.
• Individually protect electrodes before and after cleaning and sterilization to prevent damage to electrode surfaces.
• Examine handles, cables, and connector for damage or signs of wear (that is, loose cable connections, damaged pins, exposed
wires, and cable connector corrosion).
• Examine electrodes for scratched, pitted, or chipped surfaces and bubbled, scratched, or chipped epoxy coating. If any of these
conditions are found, remove the affected component from use immediately.
• Internal handles must be positioned in the sterilizer to allow recessed electrode socket (lumen) area to drain.
• Coil cables loosely away from handles during sterilization. Damage may occur if cables have tight bends or are wrapped around
the handles.
Testing:
Perform 20 J defibrillation test using a defibrillator analyzer no less than quarterly or after ten sterilization cycles, whichever comes
first.
(Continued on next page)
NOTE
:
These internal handle and electrode sterilization guidelines supersede all others for these
accessories.
WARNING:
Failure to follow the guidelines or use of the accessory beyond the limits stated may result in
damage that limits the device performance or sterilization effectiveness.
Summary of Contents for 805355
Page 1: ...Internal Defibrillation Handles and Electrodes Sterilization Guidelines ...
Page 2: ......
Page 4: ......
Page 39: ......