• Periodically inspect electrical cords and cables for damage or signs of wear. Discontinue use and replace if damaged.
• To avoid electrical shock, always unplug the power cord from the wall outlet before cleaning the device. DO NOT immerse the
device in any fluids.
• If the device is used by multiple persons (such as rental devices), a low-resistance, main flow bacteria filter should be installed in-
line between the device and the circuit tubing to prevent contamination.
• Be sure to route the power cord to the outlet in a way that will prevent the cord from being tripped over or interfered with by
chairs or other furniture.
• This device is activated when the power cord is connected.
• For safe operation when using a humidifier, the humidifier must always be positioned below the breathing circuit connection at
the mask. The humidifier must be level for proper operation.
Note:
Please see the “Limited Warranty” section of this manual for information on warranty coverage.
Cautions
A Caution indicates the possibility of damage to the device.
• Medical electrical equipment needs special precautions regarding EMC and needs to be installed according to EMC information.
Contact your home care provider regarding EMC installation information.
• Mobile RF communications equipment can affect medical electrical equipment.
• Pins of connectors marked with the ESD warning symbol shall not be touched and connections shall not be made without
special precautions. Precautionary procedures include methods to prevent build-up of electrostatic charge (e.g., air conditioning,
humidification, conductive floor coverings, non-synthetic clothing), discharging one’s body to the frame of the equipment or
system or to earth. It is recommended that all individuals that will handle this device understand these precautionary procedures
at a minimum as part of their training.
• Before operating the device, ensure that the SD card/filter access door and the modem access door are both closed whenever any
of the accessories such as the Link Module or Modem are not installed. Refer to the instructions that came with your accessory.
• Condensation may damage the device. If this device has been exposed to either very hot or very cold temperatures, allow it to
adjust to room temperature (operating temperature) before starting therapy. Do not operate the device outside of the operating
temperature range shown in the Specifications.
• Do not use extension cords with this device.
• Make sure the filter area on the side of the device is not blocked by bedding, curtains, or other items. Air must flow freely around
the device for the system to work properly.
• Do not place the device directly onto carpet, fabric, or other flammable materials.
• Do not place the device in or on any container that can collect or hold water.
• A properly installed, undamaged Philips Respironics blue pollen filter is required for proper operation.
• Tobacco smoke may cause tar build-up within the device, which may result in the device malfunctioning.
• Dirty inlet filters may cause high operating temperatures that may affect device performance. Regularly examine the inlet filters as
needed for integrity and cleanliness.
• Never install a wet filter into the device. You must ensure sufficient drying time for the cleaned filter.
• Always ensure that the DC power cord securely fits into your therapy device prior to use. Contact your home care provider or
Philips Respironics to determine if you have the appropriate DC cord for your specific therapy device.
• When DC power is obtained from a vehicle battery, the device should not be used while the vehicle’s engine is running. Damage
to the device may occur.
• Only use a Philips Respironics DC Power Cord and Battery Adapter Cable. Use of any other system may cause damage to the device.
Contraindications
When assessing the relative risks and benefits of using this equipment, the clinician should understand that this device can deliver
pressures up to 20 cm H
2
O. In the event of certain fault conditions, a maximum pressure of 40 cm H
2
O is possible. Studies have
shown that the following pre-existing conditions may contraindicate the use of CPAP therapy for some patients:
• Bullous Lung Disease
• Pathologically Low Blood Pressure
• Bypassed Upper Airway
• Pneumothorax
• Pneumocephalus has been reported in a patient using nasal Continuous Positive Airway Pressure. Caution should be used when
prescribing CPAP for susceptible patients such as those with: cerebral spinal fluid (CSF) leaks, abnormalities of the cribriform
plate, prior history of head trauma, and/or pneumocephalus. (Chest 1989; 96:1425-1426)
The use of positive airway pressure therapy may be temporarily contraindicated if you exhibit signs of a sinus or middle ear
infection. Not for use with patients whose upper airways are bypassed. Contact your health care professional if you have any
questions concerning your therapy.