Device Alerts
Device alerts are pop-ups that show up on the UI screen. There are 5 types of alerts described here:
• Status:
These alerts are just the pop-up screen.
• Notification:
These alerts consist of the pop-up screen in addition to a blinking Power LED on top of the device.
• Alert 1:
These alerts consist of the pop-up screen, a blinking Power LED and an audible beep when displayed. This
alert will not occur during therapy.
• Alert 2:
These alerts consist of the pop-up screen, a blinking Power LED and an audible beep when displayed. This
alert can occur during therapy.
• Safe State:
These alerts consist of the pop-up screen, a blinking Power LED and a repeating audible beep.
Note:
Status alerts automatically time out after 30 seconds and their pop up screens disappear. All other alerts
must be acknowledged to clear.
Alert Summary Table:
The following table summarizes the alerts.
Alert
Icon
Type
Description
Possible Cause
Action
Data Activity: Do
not remove SD
card.
Status
SD card read/write
underway.
n/a
No action needed.
Change Accepted
Status
Confirms acceptance
of prescription change
or device upgrade.
n/a
No action needed.
EZ-Start
Pressure
Incremented to
xx.x
Status
Displays when
EZ-Start mode is
enabled and device
is increasing therapy
pressure setting for
the next session.
n/a
No action needed.
Oximetry: Good
Connection
(icon only)
Status
Displays on the
therapy screen when
the blower is on
and 3 seconds of
good connection is
detected. Appears
at the beginning of
therapy. This screen
will not display
again if the finger
probe is removed
and reapplied unless
therapy is stopped
and restarted.
n/a
No action needed.
Pair?: 123456
Yes/No
Status
Prompts to accept or
decline pairing to a
Bluetooth compatible
device. This device
can be identified by
the digits displayed.
n/a
Rotate control dial to
accept pairing (Yes),
or decline (No), then
press control dial to
confirm selection.