Problem
Possible cause
Solution
The Philips
BlueControl
suddenly
interrupts the
treatment and
slowly flashes blue.
The device
measured a
too high skin
temperature
and paused the
treatment to allow
the skin to cool
down.
No action required.
The treatment restarts
automatically after the device
measures a skin temperature
in the safe range.
The Philips
BlueControl is
switched off with
a long beep and
the LED indicator
flashes orange.
The Philips
BlueControl was
interrupted too
many times during
treatment, for
instance because
the sensor sensed
a too high skin
temperature or
because maximum
treatment time of
40 minutes was
exceeded.
If this happens a few times,
contact Philips.
The LED indicator
lights up solid
orange when
I press the on/off
button.
The device
temperature
is too high
and the Philips
BlueControl
switched off for
safety reasons.
Let the Philips BlueControl
cool down for 10 minutes.
Put the Philips BlueControl
back onto your skin and press
the on/off button to resume
treatment.
ENGLISH
22
Summary of Contents for BlueControl PSD1211
Page 1: ...PSD1211 ...
Page 2: ......
Page 3: ...1 2 3 4 5 6 7 13 10 9 11 12 8 1 ...
Page 4: ...PSD1211 ENGLISH 6 DANSK 32 SUOMI 58 NORSK 84 SVENSKA 110 ...
Page 5: ...PSD1211 ENGLISH 6 DEUTSCH 32 FRANÇAIS 61 ITALIANO 89 NEDERLANDS 117 ...
Page 31: ...ENGLISH 31 This symbol means Read the user manual This symbol indicates the batch code ...
Page 82: ...SUOMI 82 Ei valkaisua Ei rumpukuivausta Ei silitystä Ei kuivapesua Valmistettu Kiinassa ...
Page 83: ...SUOMI 83 Lue käyttöopas Eränumero ...
Page 135: ...SVENSKA 135 Den här symbolen betyder Läs användarhandboken Den här symbolen anger satskoden ...
Page 161: ...ENGLISH 31 This symbol means Read the user manual This symbol indicates the batch code ...
Page 276: ...3227 000 2386 1 ...