UV LAMP
Lighting the UV lamp for 24 hours
Follow the procedure below when lighting UV lamp for 24 hours.
Before sterilization by UV lamp, remove all attachments in the chamber (tray, side support, rear duct, fan,
humidifying pan, humidifying pan cover, gas injection nozzle, gas injection nozzle tube) and clean the
interior, inner door and UV lamp thoroughly.
Description of operation
Key operated
Indication after operation
1
Press calibration key
for 5 seconds.
CAL
The left digit in the digital
temperature indicator blinks.
2
By pressing numerical value shift
key and digit shift key,
set the figure to F02.
When pressed, the figure of settable digit changes.
When pressed, the changeable digit
is shifted.
3
Press enter key.
ENT
“000” is displayed in the digital
CO
2
density indicator. (unit: min)
4
By pressing numerical value shift
key, set the figure to 001.
When pressed, the figure of
settable digit changes.
5
Press enter key.
ENT
Set value is memorized and the display returns to
normal display mode.
●
This setting should be executed with the outer door closed and the UV lamp off.
●
The UV lamp keeps on for 24 hours after the setting. The setting is canceled when the outer door is
opened. Follow the above procedure to set the UV 24-hour mode again.
●
The unit keeps running with set temperature, CO
2
density and O
2
density during UV lighting.
Pressing calibration key (CAL) for about 5 seconds leads the calibration mode. In the calibration mode,
the calibration of temperature, CO
2
density and O
2
density is possible. Wrong key operation affects the
basic performance. After about 90 seconds, the unit returns to chamber temperature display mode
automatically. In this case, the set value that has not been fixed pushing enter key (ENT) is not
changed.