22
REMOVING INNER ATTACHMENTS
CAUTION
Wear rubber gloves when performing maintenance on the chamber.
Failure to wear gloves may
result in cuts or abrasions from sharp edges or corners.
Be careful not to damage the water level sensor or the UV lamp in the rear duct. (When an optional UV
system set MCO-19UVS is installed.)
Do not use detergents or antiseptic solutions with acid, alkali, or chlorine. Doing so may cause
discoloration, corrosion, or rusting.
1.
Turn OFF the power switch of the incubator.
2.
Open the outer and inner doors and pull out the trays. (Fig. 1)
3.
Lift the humidifying pan cover off from the pins at the rear side. (Fig. 2)
4.
Pull out the humidifying pan. (Fig. 3)
5.
Remove the screw by securing the clamp for the side support. (Fig. 4)
6.
Remove the clamp. (Fig. 5)
7.
Lift the side support off of the pins. (Fig. 6)
Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6