35
Precautions when using the UV lamp
The cautions below are applicable when an optional UV system set MCO-19UVS is installed.
●
Always use humidifying pan and humidifying pan cover
The humidifying pan and humidifying pan cover prevent the UV light from escaping. Make sure they are
installed even if you do not need humidity.
●
Notice of recommended replacement of UV lamp
This unit is provided with a function to notify the recommendation of UV lamp replacement when the
accumulated ON time of UV lamp is over about 1,000 hours. The blink of the UV indicator on the control
panel recommends the replacement of UV lamp. For the replacement, contact our sales representative
or agent.
●
Location of UV lamp
The UV lamp is OFF when the outer door is opened as it is operated to outer door switch. Check the UV
lamp is ON with the inner door closed once a month by activating the outer door switch. If the lamp is
not ON, the lamp needs to be replaced. Contact our sales representative or agent.
The UV lamp is located in the rear duct. Take care not to damage the lamp at the time of
installation/removal of attachments or humidifying pan.
WARNING
Never expose the eyes directly to UV light
as UV light can cause permanent damage to eyes. Never
remove humidifying pan cover when UV light is ON (When an optional UV system set MCO-19UVS is
installed).
Hazardous UV light.
Do not press door switch.