12
INCUBATOR COMPONENTS
12. Set key
(SET): Pressing this key to enter the setting mode, and the digits to be set will blink.
13. Calibration key
(CAL): By pressing this key for approximately 5 seconds, the unit enters calibration
function mode. See page 39 for the details. Also, used to change the UV lamp ON period. See page
34 for the details.
14. Alarm buzzer stop key
(BUZZER): Press this key to silence the buzzer when the alarm operates
and the buzzer sounds.
15. N
2
/O
2
gas supply line indicator
(A/B): The lamp for the N
2
/O
2
gas
supply line currently in use lights
up. (Lamp ON; cylinder A, Lamp blinks; cylinder B)
16. CO
2
gas supply line indicator
(A/B): The lamp for the CO
2
gas supply line currently in use lights up
provided that gas auto changer MCO-5GC is installed. (Lamp ON; cylinder A, Lamp blinks; cylinder B)
17. Gas supply line switching key
(GAS SUPPLY): This is a key to select N
2
or O
2
gas supply. By
pressing this key, the N
2
or O
2
gas supply line (A/B) is changed in O
2
density set mode. The supply line
is changed automatically when N
2
or O
2
gas cylinder is empty. Similarly, by pressing this key in CO
2
density set mode, CO
2
gas supply line is changed automatically when a gas auto changer MCO-5GC
(option) is installed. The supply line is changed automatically when CO
2
gas cylinder is empty.
18. Enter key
(ENT): Pressing this key memorizes the set value in the controller.
19. Numerical value shift key
( ): Pressing this key in the setting mode causes the numerical value
to shift. In key lock mode, pressing this key makes key lock ON or OFF.
20. Digit shift key
( ): Pressing this key in the setting mode causes the changeable digit to shift.
Pressing this key more than 5 seconds enters key lock mode. See page 30 for the key lock.
12
13
14
15
18
19
20
17
16