INSTALLATION
Connecting a gas injection nozzle
In the case of control of O
2
density in the chamber, connect the gas injection nozzle to the gas injection
port by using the gas injection nozzle tube enclosed (inner diameter; 5 mm, outer diameter; 9 mm, length;
300 mm). This helps faster recovery of humidity after opening of the inner door.
Fill the humidifying pan with sterile distilled water so that the gas injection nozzle can be under the water
level.
1.
Inside of unit is arranged as shown in Fig. 1 when humidifying pan cover removed. There is a gas
injection port in the lower left-hand corner.
2.
Connect gas injection nozzle to gas injection port using gas injection nozzle tube. (Fig. 2)
3.
Connecting gas injection nozzle is as shown in Fig. 3.
Fig.1 Fig.2 Fig.3
Gas injection nozzle
Gas injection port
Gas injection nozzle tube