10
INCUBATOR COMPONENTS
1.
Outer door:
Sticks to frame with magnetic seal. Door heater is installed in the door panel. The
door opening is reversible. Contact our sales representative or agent to change the door hinge from left
to right or vice versa.
2. Inner door:
Made of tempered glass, however avoid excessive impact on the glass.
3. Leveling feet:
Screw type for adjusting the height. Adjust the feet so that the unit can be level.
4. Tray:
Can be pulled toward you.
5. Side support:
Right and left side supports can be removed for disinfection. See page 22 and 23.
6. Rear duct:
Flow path for circulating air. Removable.
7. Fan (inside the rear duct):
Made from polypropylene resin. Can be sterilized by an autoclave.
8. Sample air outlet:
This also functions as an internal gas outlet. Use only a supplied cap(9).
9.
Sample air outlet cap:
Always attach this cap except at the time of using of sample air outlet.
10. Connecting port A/B for CO
2
gas pipe (rear side):
When an optional component MCO-5GC (gas
auto changer) is installed, both A and B are available. If MCO-5GC is not installed, only A is available.
Refer to page 16 for gas cylinder connection. Ensure that the gas pressure is set at 0.03 MPa(G) (0.3
kgf/cm
2
(G), 4.4 psi(G)). Refer to page 31 for automatic gas cylinder changeover.
11. Connecting port A/B for N
2
/O
2
gas pipe (rear side):
Refer to page 17 for gas cylinder connection.
Ensure that gas pressure is set at 0.05 MPa(G) (0.5 kgf/cm
2
(G), 7.3 psi(G)). Refer to page 31 for
automatic gas cylinder changeover.
12. Door switch:
Detects the door opening/closing and stops the circulating fan and electromagnetic
valve for CO
2
and N
2
/O
2
when door is open. UV lamp is also deactivated by door opening (When an
optional UV system set MCO-19UVS is installed).
13. Humidifying pan:
Use sterile distilled water to fill the humidifying pan.
14. Humidifying pan cover:
Prevents UV light being exposed to the chamber. When filling the
humidifying pan, lift the front side and take out the humidifying pan. See page 25 for details.
15. UV lamp
(When an optional UV system set MCO-19UVS is installed)
:
This UV lamp does not generate ozone. Never look at the UV light directly. For replacement, contact
our sales representative or agent.
16. Water level sensor:
Detects the water level in the humidifying pan. See page 26 for details.
17. Remote alarm terminal:
Refer to page 13.
18. Access port:
When not in use, cap with the silicon caps on both outside and inside.
19. Power switch:
Main switch of the unit (ON-“I”, OFF-“O”). Also functions as an over-current
breaker. The switch is covered by a switch cover to prevent the accidental push. To turn on or off the
switch, remove the switch cover by loosening the screw. See figure on the right.
20. Glow starter
(When an optional UV system set MCO-19UVS is installed)
:
For UV lamp (model;
FG-7P)
21. Power supply removable cord
22. Handle:
Outer door handle. When moving the handle from upper to a lower position remove it from
original position first, then remove two cups and screws at the lower position and attach the handle there.
23. Gas injection nozzle:
Refer to page 18.