Phcbi MCO-18AC Series Operating Instructions Manual Download Page 34

32 

 

   ALARMS, SAFETY, AND SELF-DIAGNOSIS 

 
The incubator supports the following alarms, safety functions, and self-diagnostic functions. 

Alarms, safety and self-diagnosis for culture operations 

Alarm or safety 

function 

Conditions Indication 

Buzzer 

Safety 

operation 

High limit 

temperature alarm 

If the chamber temperature exceeds 

the high limit temperature alarm set 

value. 

Over heat lamp lights. 

E12 or E16 and chamber 

temperature are displayed 

alternately. 

Continuous tone 

Heater OFF 

Remote alarm 

Automatic set 

temperature alarm 

If the chamber temperature deviates 

from the set temperature by ±1 

o

C or 

more. 

All digits on the digital 

temperature indicator blink. 

Intermittent tone with 

15 minutes delay. 

Remote alarm with 15 

minutes delay 

Automatic set CO

density alarm 

If the chamber CO

2

 density deviates 

from the set value by ±1 % or more. 

All digits on the digital CO

density indicator blink. 

Intermittent tone with 

15 minutes delay. 

Remote alarm with 15 

minutes delay 

Door check 

The outer door is open. 

Door lamp lights 

----- 

Fan motor OFF 

CO

2

 valve is closed 

UV lamp OFF* 

(60 seconds after) 

Heater OFF 

Auto-return 

When there is no key pressing in 

each setting mode for 90 seconds. 

Normal display mode. 

----- 

The setting mode is 

canceled. 

Key lock 

When the key lock is “ON”. 

----- 

----- 

The setting is disabled. 

CO

2

 gas cylinder 

empty 

If the CO

2

 density does not increase 

when the gas valve is opened. 

E01 is displayed alternately 

with the temperature on the 

digital temperature 

indicator. 

Intermittent tone 

Remote alarm 

Gas line changeover  When the gas supply line is switched. 

(only when MCO-21GC is installed) 

E01 is displayed alternately 

with the temperature on the 

digital temperature 

indicator. 

CO

2

 gas supply line 

indicator of empty blinks. 

Intermittent tone 

Gas supply line is 

altered. 

Remote alarm 

Chamber 

temperature sensor 

abnormality 

If the temperature sensor is 

disconnected. 

E05 is displayed alternately 

with the temperature on the 

digital temperature 

indicator. 

Intermittent tone 

Heater OFF 

Remote alarm 

If the temperature sensor is short 

circuited. 

E06 is displayed alternately 

with the temperature on the 

digital temperature 

indicator. 

Ambient temperature 

sensor abnormality 

If the ambient temperature sensor is 

disconnected. 

E09 is displayed alternately 

with the temperature on the 

digital temperature 

indicator. 

Intermittent tone 

Remote alarm 

If the ambient temperature sensor is 

short circuited. 

E10 is displayed alternately 

with the temperature on the 

digital temperature 

indicator. 

CO

2

 sensor 

abnormality 

If the output voltage of the CO

2

 sensor 

is abnormal. 

E11 is displayed alternately 

with the temperature on the 

digital temperature 

indicator. 

Intermittent tone 

CO

2

 valve is closed 

Remote alarm   

Main heater 

abnormality 

If the high limit temperature alarm 

alarm operates, or if the main heater is 

open circuit, or the main heater relay 

is short circuited. 

E12 is displayed alternately 

with the temperature on the 

digital temperature 

indicator. 

Intermittent tone 

Remote alarm 

Bottom heater 

abnormality 

If the bottom heater goes open circuit, 

or the bottom heater relay is short 

circuited. 

E13 is displayed alternately 

with the temperature on the 

digital temperature 

indicator. 

Intermittent tone 

Remote alarm 

Door heater 

abnormality 

If the door heater goes open circuit, 

or the door heater relay is short 

circuited. 

E14 is displayed alternately 

with the temperature on the 

digital temperature 

indicator. 

Intermittent tone 

Remote alarm 

Disconnection of 

sensor for each 

heater 

If the relay of main heater, bottom 

heater or sensor box heater goes open 

circuit. 

E16 is displayed alternately 

with the temperature on the 

digital temperature 

indicator. 

Intermittent tone 

Remote alarm 

Low humidifying 

water* 

If the water in the humidifying pan is 

about 0.8 L. 

Water level alarm lamp 

(RH PAN) blinks. 

----- ----- 

UV lamp failure* 

When the UV lamp is burned out. 

E18 is displayed alternately 

with the temperature on the 

digital temperature 

indicator. 

Intermittent tone 

Remote alarm 

Recommendation of 

new UV lamp* 

The accumulated ON time is over about 

1,000 hours. 

UV indicator blinks   

when the UV lamp is OFF. 

----- ----- 

*When an optional UV system set MCO-18UVS3 is installed. 

 

• The alarm can be canceled by pressing the alarm buzzer stop key (BUZZER), but the remote alarm cannot be 

stopped.    And the high limit temperature alarm cannot be silenced with the alarm buzzer stop key (BUZZER). 

• E01 is cleared automatically when the gas is connected correctly and the buzzer is silenced with the alarm buzzer 

stop key (BUZZER).    When MCO-21GC is installed, press the alarm buzzer stop key (BUZZER) to silence the alarm 

after changeover of gas supply line. 

• If one of E05 to E18 (except for E12, E13, and E14) is displayed, contact our sales representative or agent. 

Summary of Contents for MCO-18AC Series

Page 1: ...AC Series Please read the operating instructions carefully before using this product and save the operating instructions for future use See page 46 for all model numbers Operating Instructions CO2 Inc...

Page 2: ......

Page 3: ...YING PAN 21 WATER LEVEL SENSOR MCO 18AC with MCO 18UVS3 23 ROUTINE MAINTENANCE 24 CORRECT OPERATION 24 SETTING CHAMBER TEMPERATURE AND CO2 DENSITY 25 SETTING OF ALARM RESUME TIME 26 Operation after po...

Page 4: ...bject to change without notice for improvement of performance or functions Contact our sales representative or agent if any page of the operating instructions is lost or the page order is incorrect Co...

Page 5: ...lure to observe CAUTION signs could result in injury to personnel and damage to the unit and associated property Symbol shows This symbol means caution This symbol means an action is prohibited This s...

Page 6: ...wer supply outlet is not grounded it will be necessary to install a ground by qualified engineers Never ground the unit through a gas pipe water main telephone line or lightning rod Such grounding may...

Page 7: ...ric shock or fire When removing the power supply plug from the power supply outlet grip the power supply plug not the cord Pulling the cord may result in electric shock or fire by short circuit Discon...

Page 8: ...ces such as acid or alkali in this unit if the container cannot be sealed These may cause corrosion of inner components or electric parts Check the setting when starting up of operation after power fa...

Page 9: ...ce This symbol indicates an earth This symbol means ON for a power switch This symbol means OFF for a power switch ENVIRONMENTAL CONDITIONS This equipment is designed to be safe at least under the fol...

Page 10: ...3 4 5 6 7 9 inside 10 11 13 19 20 inside 14 15 Rear Right Side When an optional UV system set MCO 18UVS3 is installed 12 16 21 Rear Left Side When an optional UV system set MCO 18UVS3 is installed 18...

Page 11: ...1GC is not used only A is available Refer to page 14 for gas cylinder connection Ensure that the gas pressure is set at 0 03 MPa G 0 3 kgf cm2 G 4 4 psi G Refer to page 41 for gas auto changer 13 Door...

Page 12: ...s lamp lights when the chamber temperature reaches the high limit temperature 8 CO2 inject lamp INJECT This lamp lights when CO2 gas is being injected 9 Digital CO2 density indicator CO2 Normally this...

Page 13: ...de causes the numerical value to shift In key lock mode pressing this key makes key lock ON or OFF See page 27 for details 17 Digit shift key Pressing this key in the setting mode causes the changeabl...

Page 14: ...he floor is uneven or the installation is not level the incubator will be unstable and this may cause accident or injury To avoid vibration and noise always make sure that the installation is stable A...

Page 15: ...ith qualified service personnel WARNING To prevent electric shock always ground the incubator If grounding is not possible then have a ground installed by qualified service personnel If the incubator...

Page 16: ...ve will operate and there may an explosive sound When the MCO 21GC is not mounted Using the gas tube that is provided connect the gas regulator for the CO2 gas cylinder to connecting port A for CO2 ga...

Page 17: ...water spills from the humidifying pan or if the doors are left open for a long time condensation may form on the inside of the doors If that occurs wipe off the condensation with a dry sterile gauze I...

Page 18: ...en closing the outer door Use the handle when closing the outer door Holding the door in other places may cause injury by getting fingers caught in the door Do not lean on the outer door Doing so may...

Page 19: ...on corrosion or rusting Be careful to keep the diluted detergent or water out of the temperature sensor the CO2 gas injection hole and the fan motor shaft bearing Fig 2 Also do not wash the temperatur...

Page 20: ...orrosion or rusting 1 Turn OFF the power switch of the incubator 2 Open the outer and inner doors and pull out all the trays Fig 1 3 Lift the front of the tray supports and pull them out Fig 2 4 When...

Page 21: ...r the side support Fig 8 10 Remove the clamp Fig 9 11 Lift the side supports off of the pins Fig 10 12 Remove the silicon caps of the access port one each from interior Fig 11 and exterior Fig 12 1 Pu...

Page 22: ...failure or insufficient decontamination 3 To install the duct confirm 4 pins are securely installed in the 4 holes of duct Fig 2 CAUTION If the duct is fixed insecurely the intended velocity performan...

Page 23: ...low temperature water will lower the temperature and humidity in the chamber CAUTION Set the humidifying pan lengthwise at the center of chamber width At that time push the humidifying pan all the way...

Page 24: ...cessive force to the sensor lead wire Note The humidity in the incubator chamber is adjusted to the optimum setting To prevent condensation on the surface inside chamber and the inner door there is a...

Page 25: ...ing pan The water level alarm lamp RH PAN blinks if the water level sensor does not come down completely If necessary set the humidifying pan again in the proper location The water level sensor detect...

Page 26: ...14 2 Remove the packing materials from the chamber and inner attachments Clean and disinfect the chamber and all the inner attachments referring to above ROUTINE MAINTENANCE 3 Add approximately 1 5 L...

Page 27: ...figure to 33 0 When pressed the changeable digit is shifted When pressed the figure of settable digit changes 4 Press enter key ENT Set temperature is memorized Left digit in the digital CO2 density i...

Page 28: ...digit changes 6 Press enter key ENT The setting is memorized and the current chamber temperature is displayed The settable alarm resume time are 0 10 20 30 40 50 or 60 minutes The setting is 000 010 0...

Page 29: ...etting change from key lock OFF to key lock ON Description of operation Key operated Indication after operation 1 The current chamber temperature and CO2 density are displayed 2 Press digit shift key...

Page 30: ...door is closed The default setting is for 5 minutes The period can be changed when necessary as shown in the page 30 If the outer door is not opened for at least 12 consecutive hours the UV lamp will...

Page 31: ...an be confirmed from the front of the humidifying pan cover UV light is harmful to eyes so do not light the UV lamp when the inner door or the humidifying pan cover is open WARNING Never expose eyes d...

Page 32: ...t value is memorized and the display return to normal display mode The UV timer can be set from 0 minute to 30 minutes The default setting is for 5 minutes When the UV timer is set to zero the UV lamp...

Page 33: ...ue shift key set the figure to 001 When pressed the figure of settable digit changes 5 Press enter key ENT Set value is memorized and the display returns to normal display mode This procedure must be...

Page 34: ...e temperature on the digital temperature indicator Intermittent tone Remote alarm If the ambient temperature sensor is short circuited E10 is displayed alternately with the temperature on the digital...

Page 35: ...bration key CAL 2 Press the calibration key CAL for about 5 seconds 3 The 3 on the digital temperature indicator blinks and the digital CO2 density indicator goes out 4 Adjust the set value to the act...

Page 36: ...n the digital CO2 density indicator blinks and the temperature indicator goes out 9 Set the present correct CO2 density with the digit shift key and numerical value shift key then press the enter key...

Page 37: ...a week The primary pressure of less than 3 8 MPa G 38 kgf cm2 G 551 psi G means a little gas in the cylinder Replace the cylinder soon It is used near an electric product which emits an electromagnet...

Page 38: ...ecomes stable enough approx 1 hour after turning ON the power switch or recovering from power failure Note If the problem still has not been solved after trying the above checks and countermeasures or...

Page 39: ...n und elektronischen Ger ten private Haushalte Dieses Symbol auf Produkten und oder begleitenden Dokumenten bedeutet dass verbrauchte elektrische und elektronische Produkte nicht mit gew hnlichem Haus...

Page 40: ...ter les autorit s locales ou votre revendeur afin de conna tre la proc dure d limination suivre Spanish Informaci n sobre la eliminaci n para los usuarios de equipos el ctricos y electr nicos usados p...

Page 41: ...orma es sobre o m todo de elimina o correcto Italian Informazioni per gli utenti sullo smaltimento di apparecchiature elettriche ed elettroniche obsolete per i nuclei familiari privati Questo simbolo...

Page 42: ...act op met de lokale overheid of uw leverancier en vraag wat de juiste verwijderingsmethode is Swedish Information om kassering f r anv ndare av elektrisk elektronisk utrustning privata konsumenter Om...

Page 43: ...pen the valve of both gas cylinders 4 The CO2 gas supply line indicator of CO2 gas supply line currently used is ON Only when MCO 21GC is installed CO2 gas supply line indicator A CO2 gas supply line...

Page 44: ...s quantity in the cylinder before disconnecting it Following shows the procedure to change the CO2 gas supply line manually Only when MCO 21GC is installed Example Change of CO2 gas cylinder from B to...

Page 45: ...the bottom and attach the protective stickers to the 4 corners on the top surface of the incubator that is to be on the bottom Fig A 4 Secure stacking plates A with the screws that are provided 5 Remo...

Page 46: ...44 STACKING INCUBATORS Fig A Front panel Stacking plate A Protective sticker Hook Stacking plate B Front side...

Page 47: ...sity display Digital display Air circulation Fan assisted Air filter 0 3 m Efficiency 99 97 or more Alarms Automatic set temperature alarm Automatic set CO2 density alarm High limit temperature alarm...

Page 48: ...variation 0 15 ambient temperature 25 o C setting 37 o C CO2 5 no load Chamber humidity 95 R H 5 R H Heat emission Max 1116 kJ h Applicable environment condition Temperature 5 o C to 35 o C Humidity...

Page 49: ...n of the unit Unit interior No contamination Decontaminated Contaminated Others 3 Instructions for safe repair maintenance disposal of the unit a The unit is safe to work on b There is some danger see...

Page 50: ......

Page 51: ......

Page 52: ...Printed in Japan LDCL039500 0 S0418 0 1 1 1 Sakada Oizumi machi Ora gun Gunma 370 0596 Japan PHC Corporation 2018...

Reviews: