![Phcbi MCO-18AC Series Operating Instructions Manual Download Page 30](http://html1.mh-extra.com/html/phcbi/mco-18ac-series/mco-18ac-series_operating-instructions-manual_1536743030.webp)
28
UV LAMP (MCO-18AC with MCO-18UVS3)
The clauses below are applicable when an optional UV system set MCO-18UVS3 is installed.
The UV lamp is located inside the duct to sterilize the water in the humidifying pan and air circulating in
the chamber.
When all the inner attachments are installed correctly, only the inside of the duct and the inside of the
humidifying pan cover are exposed to UV light.
Correctly install all of the inner attachments when starting a cell culture, and never turn ON the UV
lamp when the humidifying pan cover is removed.
Always install the humidifying pan cover even when using the incubator without turning ON the UV
lamp. Leaving the cover uninstalled will affect the chamber temperature distribution and the humidity
recovery performance.
The UV lamp stays lit for a preset period of time after the outer door is closed. The default setting is for
5 minutes. The period can be changed when necessary as shown in the page 30.
If the outer door is not opened for at least 12 consecutive hours, the UV lamp will light for the preset
time period every 12 hours. The period can be changed when necessary as shown in the page 30.
The recommended replacement time for the UV lamp (i.e., when the UV output ratio drops to approx.
70% of its initial value) is when the accumulated ON time reaches 1,000 hours. When the accumulated
ON time reaches approximately 1,000 hours, the UV indicator blinks (when the UV lamp is not lit). It is
recommended that the UV lamp be quickly replaced at this point. Contact our sales representative or
agent for information on replacing the UV lamp.
If the UV lamp burns out, E18 is displayed on the digital temperature indicator. If this occurs, replace
the UV lamp. When replacing the UV lamp, replace the glow starter (type FG-7P) at the same time.
Contact our sales representative or agent for information on replacing the UV lamp.