INSTALLATION
1. Remove the packing tape and clean up.
Remove all the tapes that are securing the doors and inner attachments. Open the doors for ventilation. If
the outer panels are dirty, dampen a cloth with a diluted neutral detergent and wipe them. (Using an
undiluted solution may damage the plastic. Follow the diluting instructions for the detergent that is used.)
Wipe off the residual detergent with a wet cloth and then wipe off any moisture.
Note:
Remove the cable tie banding the power supply cord. Prolonged banding may cause the corrosion of
the cord coating.
WARNING
Do not leave plastic wrapping bags within reach of children.
If a bag is placed over a child’s head, it
can block mouth and nose and cause suffocation.
2. Adjust the leveling feet.
Adjust the leveling feet by turning them counterclockwise to level
the incubator. (See the illustration on the right.)
3. Ground the incubator.
Ground the incubator during installation to prevent electric shock
in case the insulation is not sufficient. If there is no ground wire at
the location, contact with qualified service personnel.
When a ground must be installed
If a grounded 3-pole outlet is not available, then a ground must
be installed. Contact with qualified service personnel.
WARNING
To prevent electric shock,
always ground the incubator.
If grounding is not possible, then have a ground
installed by qualified service personnel. If the incubator is not grounded, it may result in electric shock.
Never connect the ground wire to a gas pipe, water pipe, lightning rod, or telephone ground wire.
Doing so may cause electric shock
.
Installing a ground fault circuit breaker
If using the incubator in a location with moisture or humidity cannot be avoided, then it is recommended
that a ground fault circuit breaker be installed in the power supply circuit (i.e., the power supply at the
incubator). Have the circuit breaker installed by qualified service personnel.
CAUTION
Do not climb on the incubator or place objects on top of it.
Doing so may damage it or cause it to fall
over, resulting in injury. If it is to be stacked, refer to page 43 and stack it securely.
When the incubator is not in use
Empty the water from the humidifying pan and remove moisture from the chamber. Make sure that the
chamber is completely dry before closing the doors. Failure to do so may result in damage.
Before moving the incubator
Before moving the incubator, empty the water from the humidifying pan, disconnect the power supply plug
from the outlet, and make sure that the power supply cord will not be damaged. Failure to do so may
result in electric shock or fire.
Leveling feet