24
ROUTINE MAINTENANCE
To use this unit in a clean condition, clean the chamber and all the inner attachments at least once a
month.
1.
Remove all the inner attachments by the procedures shown on page 18.
2.
Clean the chamber and all the inner attachments by the procedures shown on page 17.
3.
Install all the inner attachments by the procedures shown on page 20.
●
When there is excessive dirt, contact our sales representative or agent.
CORRECT OPERATION
Use the following procedure to start trial operation or actual operation of the incubator.
1.
Install the incubator correctly, referring to “INSTALLATION
”
on page 14.
2.
Remove the packing materials from the chamber and inner attachments. Clean and disinfect the
chamber and all the inner attachments, referring to above “ROUTINE MAINTENANCE”.
3.
Add approximately 1.5 L of sterile distilled water to the humidifying pan. (Refer to page 21.)
4.
Connect the power supply cord to the outlet. Turn ON the power switch on the rear left side of the
incubator.
Note:
The humidity in the incubator chamber is adjusted to the optimum setting. To prevent condensation on the
surface inside chamber and the inner door, there is a low-temperature area under the humidifying pan in
the bottom of chamber to recondense evaporated moisture. Condensation may occur around the
humidifying pan at the bottom of the chamber (on the inside of the humidifying pan cover), but this does
not indicate a problem.