REMOVING INNER ATTACHMENTS
CAUTION
Wear rubber gloves when performing maintenance on the chamber.
Failure to wear gloves may
result in cuts or abrasions from sharp edges or corners.
Be careful not to damage the water level sensor or the UV lamp in the duct (if an optional UV system set
MCO-18UVS3 is installed.).
Do not use detergents or antiseptic solutions with acid, alkali, or chlorine. Doing so may cause
discoloration, corrosion, or rusting.
1.
Turn OFF the power switch of the incubator.
2.
Open the outer and inner doors and pull out all the trays. (Fig. 1)
3.
Lift the front of the tray supports and pull them out. (Fig. 2)
4.
(When an optional UV system set MCO-18UVS3 is installed) Lift the humidifying pan cover off from the
pins at the rear side. (Fig. 3)
5.
Pull out the humidifying pan. (Fig. 4)
6.
Loosen the two screws securing the fan cover and take off the fan cover. (Fig. 5)
7.
Lift the duct and remove it from the pins at the rear side. (Fig. 6)
Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Tray support
Pins
Humidifying pan cover
Humidifying pan
Fan cover
Duct
Screws