29
Precautions when using UV lamp
The clauses below are applicable when an optional UV system set MCO-18UVS3 is installed.
Always use the humidifying pan and humidifying pan cover.
The humidifying pan and humidifying pan cover prevent UV light from escaping. Always use them even
when not humidifying. To check whether the UV lamp is lit, open the outer door and then press the door
switch with the inner door still closed. Visible blue light can be confirmed from the front of the humidifying
pan cover. UV light is harmful to eyes, so do not light the UV lamp when the inner door or the humidifying
pan cover is open.
WARNING
Never expose eyes directly to UV light
as UV light can cause permanent damage to eyes. Never
remove humidifying pan cover when UV light is ON (When an optional UV system set MCO-18UVS3 is
installed in MCO-18AC).
Hazardous UV light.
Do not press door switch when the inner door is opened..
Be careful when handling the UV lamp.
There is a UV lamp inside of the duct. Be careful not to damage it when installing or removing the inner
attachments or the humidifying pan.