Responses
Header character
Pump address
Respons
End character
#
ADR
yyyxxxabc
<CR>
OK
If everything is OK, or specific respons to the order sent
ERR0
Setting fault
ERR1
Order fault
ERR2
Parameter fault
ERR3
Context fault
ERR4
Checksum fault
Example of dialog
Order
#005ECHON<CR>
Response
#005OK<CR>
8.4.3 Command list
Order
Pa-
ra-
me-
ter
Description
Functions
Min
Max
De-
fault
set-
ting
ADR
xxx
Assignement
an address to
a pump in the
network.
Example:
#adrADRxxx<CR>
(only for RS-232)
● adr = pump address before order
● xxx = new address
000
255
0
DEF
none
Display of
warnings/
faults history
List of the last 10 warnings/faults stamped on the electron-
ic drive unit running time, followed by the name of the fault.
The arrow->indicates the ongoing warning/fault
Example:
*
00000:02:20 / Default list
->00000:01:37 / Default in current Yh
->00000:01:37 / PM Default
*
00000:02:20 / Warning list
->00000:02:37 / Holweck Temp. sensor
00000:01:30 / Speed controller Hall sensor
00000:00:50 / Holweck Temp. sensor
DLI
xxx
Data logger
read interval
Defines the automatic read interval of pump status (STA) in
seconds
0
255s
1s
DLR
none
Automatic
read activat-
ed
Authorizes automatic reading of pump status (STA) ac-
cording to the DLI interval (RS-232 only)
OFF
ECH
ON
or
OFF
Return char-
acter re-
ceived on the
serial link
Activated if ECHON (RS-232 only)
Deactivated if ECHOFF (RS-232 only)
OFF
IDN
none
Identification
of the prod-
uct connect-
ed with the
computer
Returns the type of pump (ATHxxxx) connected with the
computer, the software version (x, yy) followed by the re-
lease (zz)
Example: #adr,ATH500 Vx.y.zz,<CR>
Interfaces for control
52/78