8.2.3 SETUP menu
Selection
Choice
Description
Initial setting
1)
ACCESS CODE
0 to 65535
0
REMOTE CONTROL
Keyboard
Remote hard
Serial link
Profibus
EtherCAT
Choose the interface control mode.
According to ordering guide.
STAND-BY SPEED
from 15000 rpm to pump
nominal speed
Activate pump Stand-by speed (se-
lected speed) between minimum
speed and nominal speed (= speed
set at factory).
15000
BUZZER
ON
OFF
Activate the acoustic signal in case
of default. To stop the signal, press
OFF
.
OFF
THERMOSTAT
ON = 30 to 65 °C
OFF
Activate the pump
temperature set-
point
.
OFF (M version)
ON = 65 °C (MT version)
RELAY AT SPEED
-3 to 50 % of the nominal
speed
Modify the speed contact setpoint.
- 3 %
FIELDBUS PRO-
FILE
2)
0
1
Reserved for service
0
FIELDBUS AD-
DRESS
2)
RS-232 SPEED
9K6
19K2
38K4
57K6
Choose the RS-232 serial link trans-
mission speed.
9K6
RS-232 ECHO
ON
OFF
Activate all received characters to be
echoed over the RS-232 serial link.
OFF
RS SEPARATOR
0 to 255
Enter the separator character in AS-
CII: e.g. “044” for “,”.
44
RS ADDRESS
0 to 255
Number given to the pump in the se-
rial link.
0
BEARING LIVE
Warning limit
0 to 99 %
Modify the landing bearing warning
threshold.
20
NEW CODING
0 to 65535
Modify the access code.
0
1) For client's specification product, inital settings could be different (see chapter “Product concerned”).
2) Refer to the operating manual of the fieldbus interface that controls the pump.
8.2.4 SER NUM menu
Selection
Display
Initial settings
HHR
Display the HHR serial number.
XXXXXXXXXX
CONTROLLER
Display the control interface serial number of the electronic drive unit.
YYYYYYYYYY
CARTRIDGE
1)
Display the moving part serial number.
ZZZZZZZZZZ
MMCC06
1)
Display the electronic board serial number.
AAAAAAAA
1) Only if HRR connected to an ATH 500 pump.
Interfaces for control
47/78