28 . TGA 8000 Installation and Hardware Guide
NOTE: The furnace shield is shipped in the inverted position on the furnace door.
3.
Unscrew and remove the furnace shield from the furnace door, invert it then secure
it on the furnace door.
4.
Remove the top foam piece that secures the instrument in the box.
5.
Prepare to lift the instrument out of the carton.
WARNING:
Since the instrument weighs 26 kg (57 lb),
two people
are needed to
safely do this. The bench should be nearby.
DO NOT
lift the unit by the top round
dome. This dome protects the balance mechanism and is not screwed down.
AVERTISSEMENT:
Puisque l'appareil pèse 26 kg (57 lb),
deux personnes
sont
nécessaires pour le faire en toute sécurité. Le banc doit être à proximité.
NE
soulevez PAS
l'appareil par le dôme supérieur. Ce dôme protège le mécanisme
de l'équilibre et n'est pas vissé.
6.
One person should put one hand underneath the bottom of the instrument and the
other hand beneath the front side. The other person should place one hand
underneath the top of the instrument and the other hand underneath the back side.
Lift the instrument simultaneously.
7.
Place the analyzer carefully on the prepared bench.
8.
Remove the tape on top of the dome.
Gently lift the dome cover STRAIGHT UP to ensure you clear the balance mechanism.
9.
Remove the tape securing the balance and the zip tie securing the furnace.
10.
If there is an autosampler attached, cut the strap holding it in place as shown to the
right of the image below).
The procedure for unpacking the TGA is complete.
Summary of Contents for TGA 8000
Page 1: ...TGA 8000 Installation and Hardware Guide THERMAL ANALYSIS...
Page 5: ...5 Safety and Regulatory Information...
Page 19: ...19 Prepare the Laboratory...
Page 25: ...25 Installing the TGA 8000...
Page 55: ...55 TGA 8000 Hardware Maintenance...
Page 68: ...68 TGA 8000 Installation and Hardware Guide...
Page 85: ...85 TGA 8000 Autosampler...