Replacing a Thermocouple . 67
Helpful Hint: A new thermocouple is shipped in a plastic container with two caps, the cap is 10 mm
in height, place this into the furnace and raise or lower the thermocouple such that it
is just touching the bottom of the cap. Remove the cap.
5.
Plug the plastic thermocouple connector plug on the base of the thermocouple into
the base of the furnace. There are two small pins in the furnace base where the
thermocouple connects. Be careful not to bend or break the thermocouple pins in the
furnace base.