76 . TGA 8000 Installation and Hardware Guide
9.
Replace the clamp around the base of the jacket to secure it to the furnace base.
10.
Install the quartz insert
11.
Swing the red handle around and tighten the clamp.
Make sure that the red handle does not touch the lower liquid connector (Cooling
Liquid In).
12.
Attach the exhaust tube to the connector in the back of the cooling jacket.
13.
The TGA 8000 furnace plugs into the left side after you remove the left side panel.
Connecting to the Chiller
CAUTION:
An appropriate fluid must be used with the chiller. The fluid should be
effective down to –20 °C. Any lower than that would impair the conveyance
power of the chiller's pump. Distilled water is recommended for operation at
temperatures between 15 °C and 90 °C. For operation at –20 °C to 100 °C, 50%
ethylene glycol in water is recommended.
ATTENTION:
Un fluide approprié doit être utilisé avec le refroidisseur. Le fluide
doit être efficace jusqu'à -20 ° C. Tout ce qui est inférieur à celui-ci
compromettrait la puissance de transport de la pompe du refroidisseur. L'eau
distillée est recommandée pour des températures comprises entre 15 ° C et 90 °
C. Pour un fonctionnement à -20 ° C à 100 ° C, 50% d'éthylène glycol dans l'eau
est recommandé.
CAUTION:
Ensure that the cooling air vents at the front and rear of the cooler are
not blocked. Maintain a minimum space of 15–20 cm.
ATTENTION:
Assurez-vous que les évents d'air de refroidissement à l'avant et à
l'arrière du refroidisseur ne sont pas bloqués. Maintenir un espace minimal de
15-20.
Summary of Contents for TGA 8000
Page 1: ...TGA 8000 Installation and Hardware Guide THERMAL ANALYSIS...
Page 5: ...5 Safety and Regulatory Information...
Page 19: ...19 Prepare the Laboratory...
Page 25: ...25 Installing the TGA 8000...
Page 55: ...55 TGA 8000 Hardware Maintenance...
Page 68: ...68 TGA 8000 Installation and Hardware Guide...
Page 85: ...85 TGA 8000 Autosampler...