82 . TGA 8000 Installation and Hardware Guide
button
.
The Autosampler Control dialog box is displayed.
Note that the Piercing Position and Piercing Depth buttons are grayed out (inactive).
3.
Click Change to set the Piercing Mode to ON.
The Piercing Position and Piercing Depth buttons become active.
4.
Make sure that a crimped sample pan (use the Accupik volatile sample pan sealer to
crimp a volatile sample pan) is in a crucible and that the crucible is in position 1 in
the autosampler tray.
Summary of Contents for TGA 8000