ASSAYS
PerkinElmer Life and Analytical Sciences
85
Background Subtract
Mark this box to subtract background CPM from all samples. The background
value is established in one of three ways and is selected from the following:
1
st
Vial
The instrument counts the first vial in the cassette for either ten minutes or the
defined protocol count time (whichever is greater) and establishes a CPM value
for each region; these are the background values subtracted from each sample
within each region of the assay.
IPA
The instrument subtracts the background values established during the calibration
and Instrument Performance Assessment (IPA) procedures from the entire
spectrum of the samples. The background spectra are stored during these
procedures and are available for any counting region.
Manual
Enter the CPM values you would like the instrument to subtract from the entire
spectrum of the samples.
Note: In Quench Standards assays, the 1
st
Vial background
subtraction option is not available. Background subtraction is only
applied to the reported data for quench standards and has no
impact on the spectrum for each standard. Any background
subtraction that occurs in DPM assays will apply to the quench
standards used for the purpose of recalculating the quench curve
in the DPM assay.
Summary of Contents for QuantaSmart 1694267
Page 1: ...QuantaSmart For The TriCarb Liquid Scintillation Analyzer Reference Manual...
Page 9: ...v Chapter 12 Theory 197 Low Level Counting Theory 197 Alpha Beta Counting Theory 207...
Page 10: ...vi...
Page 44: ...CHAPTER 3 34 PerkinElmer Life and Analytical Sciences...
Page 118: ...CHAPTER 5 108 PerkinElmer Life and Analytical Sciences...
Page 192: ...CHAPTER 9 182 PerkinElmer Life and Analytical Sciences...
Page 198: ...CHAPTER 10 188 PerkinElmer Life and Analytical Sciences...