CHAPTER 7
130
PerkinElmer Life and Analytical Sciences
Valid # of Points
This field indicates the number of points used to generate the selected chart.
SD
This field displays the Standard Deviation of the selected IPA parameter as
calculated from all the included data points. A deleted data point is not used to
determine the Standard Deviation of the IPA parameter.
IPA Reports
The following display shows the IPA Report window with no current warning
messages. Any performance parameter that fails to meet specification will
generate a warning message at the top of the report. Any IPA or calibration
operation that fails to complete satisfactorily will result in an error message
reported for this run.
Any IPA errors or warnings displayed at the bottom of the report represent the
most recent status of any given IPA parameter that generated a message during
an IPA run. The dates for each error or warning may vary to reflect the most
recent date on which data for that individual IPA parameter was acquired.
Figure 7-3
An
IPA Report
This IPA report will print after the IPA procedure is completed.