ADVANCED FEATURES
PerkinElmer Life and Analytical Sciences
135
Chapter 8
Advanced Features
Priostat
Priostat allows you direct control over several sample analysis features. Two
different types of Priostat counting are available:
Group Priostat
This method allows you to count a set of high priority samples immediately, while
temporarily interrupting the current protocol. Group Priostat allows you to select
any existing assay for the purpose of data reduction.
To use this feature:
1. Load the samples into cassettes and attach the Priostat flag to the first
cassette to be counted.
2. Select Group Priostat from the Run menu. The Associate assay for Priostat
window is displayed.
3. Select the assay you would like to use to count the samples.
Note: The interrupted protocol number cannot be selected to count
the Group Priostat samples.
The instrument will search for the Priostat flag and begin counting the samples.
Once the Group Priostat samples are counted, the instrument automatically
resumes executing the interrupted protocol with the sample it was counting when
the Group Priostat counting commenced. The results of the Group Priostat
counting will be displayed in the Output window.
4. Print the results of the Group Priostat results by selecting the print icon in the
Output window.
Note:To end the Group Priostat protocol prior to counting all of the
samples, select End Group Priostat from the Run menu.
Summary of Contents for QuantaSmart 1694267
Page 1: ...QuantaSmart For The TriCarb Liquid Scintillation Analyzer Reference Manual...
Page 9: ...v Chapter 12 Theory 197 Low Level Counting Theory 197 Alpha Beta Counting Theory 207...
Page 10: ...vi...
Page 44: ...CHAPTER 3 34 PerkinElmer Life and Analytical Sciences...
Page 118: ...CHAPTER 5 108 PerkinElmer Life and Analytical Sciences...
Page 192: ...CHAPTER 9 182 PerkinElmer Life and Analytical Sciences...
Page 198: ...CHAPTER 10 188 PerkinElmer Life and Analytical Sciences...