PerkinElmer QuantaSmart 1694267 Reference Manual Download Page 147

 

ADVANCED FEATURES  

 PerkinElmer Life and Analytical Sciences

137 

Decay

The Decay feature allows you to confirm the existence of sample luminescence 

and determine the rate of decay of luminescence for a radioactive sample. Using 

this mode, samples are recounted until the counting procedure is manually halted. 

If the counts collected decrease with each successive count (assuming the half life 

of the nuclide is long in relation to the total count time), the radioactive sample is 

luminescent. You can view either the sample spectrum or the results plotted as a 

histogram. The histogram displays the rate at which luminescence is decaying 

from the sample. This process is useful in establishing an appropriate Pre-Count 

Delay Time for samples in an assay. 

To count samples using the Decay mode:

1. Load your sample(s) into a cassette. Attach the Priostat flag in the reset 

position to the cassette.

2. Select Decay from the Run menu. The Sample Priostat window is displayed.

Figure 8-1 Decay Window

Summary of Contents for QuantaSmart 1694267

Page 1: ...QuantaSmart For The TriCarb Liquid Scintillation Analyzer Reference Manual...

Page 2: ...tion may be reproduced in any form whatsoever or translated into any language without the prior written permission of PerkinElmer Inc Copyright 2004 PerkinElmer Inc Produced in the USA Trademarks Regi...

Page 3: ...eaner System Ventilation For adequate ventilation of this equipment a distance of 15cm must be kept from this unit and any other surfaces Electrical Safety Use proper plugs and good earth ground conne...

Page 4: ...mentation Ein Off Supply Arret alimentation Aus Caution Attention Risk of Electrical Shock Attention Risque de choc Vorsicht Spannug Gefahrliche Caution refer to accompanying documents Attention Voir...

Page 5: ...zation and Calibration 8 Steps in Performing an Assay 10 How to Perform an Assay 12 How to Create a Password 15 How to Associate an Assay to a Protocol 17 How to Count Samples 19 How to Disassociate a...

Page 6: ...w Window 38 The Instrument Status Bar 39 The Protocols Tree 40 The Replay Tree 43 Menus 44 File Menu 45 Run Menu 48 View Menu 50 Libraries Menu 51 Tools Menu 55 IPA Menu 58 Diagnostics Menu 59 Window...

Page 7: ...al Files 102 Worklist 105 Chapter 6 Libraries 109 Sample Nuclides Library 110 Quench Standards Library 113 Alpha Beta Nuclide Library 114 Alpha Beta Standards Library 116 Chapter 7 Calibration and Nor...

Page 8: ...e Maintenance 163 Storing Data 164 Operational Errors 165 IPA Errors 167 Warnings and Messages 169 System Messages 169 Chapter 10 Calculations 183 Background Correction 183 Background Threshold 183 Ch...

Page 9: ...v Chapter 12 Theory 197 Low Level Counting Theory 197 Alpha Beta Counting Theory 207...

Page 10: ...vi...

Page 11: ...r contains a serial RS 232 communications port a parallel printer port and a graphics adapter for the color monitor All RS 232 communication occurs through communications port one Communications port...

Page 12: ...trument Performance Assessment IPA IPA is an option on the TriCarb 2800TR and 2900TR Chi Square Chi Square acceptable range 7 63 to 36 19 performed as 20 0 5 minute repeat counts and tested at the 99...

Page 13: ...within your samples causing erratic results The system must be connected to a stable power source through the use of proper plugs and good earth grounding The system provides automatic recovery after...

Page 14: ...andard 133 Ba nominal less than 20 Ci for all models except the TriCarb 3170TR SL 133Ba nominal less than 1 Ci for the TriCarb 3170TR SL Observed Background Average values for Normal Count Mode Tritiu...

Page 15: ...to 12 15 20ml vials and the mini vial cassettes up to 18 6 7ml mini vials A protocol flag which you attach to the cassette identifies the protocol and assay conditions that you have defined for use w...

Page 16: ...CHAPTER 1 6 PerkinElmer Life and Analytical Sciences Sample Vials Vials and caps must conform to the following sizes Caps must not exceed the diameter of the vial Figure 1 5 Vial Sizes...

Page 17: ...rds CPM Assay DPM Single DPM Dual DPM Triple FS DPM Direct DPM Quench Standards Single Photon Counting Note DPM Dual is optional on 2800 series instruments DPM Triple is optional on 2800 and 2900 seri...

Page 18: ...n Window 2 Reset the SNC protocol flag to the reset position the flag is all the way to the left when the plug is on the left end of the cassette 3 Load the purged unquenched Carbon 14 standard into t...

Page 19: ...ers in the IPA Definition window 2 Reset the SNC protocol flag to the reset position the flag is all the way to the left when the plug is on the left end of the cassette 3 Load the purged unquenched C...

Page 20: ...CHAPTER 2 10 PerkinElmer Life and Analytical Sciences Steps in Performing an Assay 1 Select the File Open Assay menu option The Open Assay window is displayed Figure 2 2 Open Assay Window...

Page 21: ...ayed Figure 2 3 Assay Definition Window 3 In the Assay Parameters tab mark the Lock Assay box The Password field becomes enabled 4 Enter a descriptive password in the Password field 5 When you have fi...

Page 22: ...ndow 3 In each of the seven tabs of the Assay Definition window define the individual assay parameters as needed 4 Save the assay using a descriptive name 5 In the Protocols Tree of the main window se...

Page 23: ...HOWTO PerkinElmer Life and Analytical Sciences 13 Note You can also associate an assay to a flag using the File menu Figure 2 5 Protocol Tree Figure 2 6 Associate Assay Window...

Page 24: ...r an Additional Header if necessary 11 Enter an Output Data Path if you would like the data to be saved in a directory other than the default In the following example the default path is C Packard Tri...

Page 25: ...al Sciences 15 How to Create a Password 1 Select the File New Assay menu option The Select Assay Type window is displayed Figure 2 8 Select Assay Type Window 2 In the Select Assay Type window select t...

Page 26: ...ure 2 9 Assay Definition Window 3 In the Assay Parameters tab mark the Lock Assay box The Password field becomes enabled 4 Enter a descriptive password in the Password field 5 When you have completed...

Page 27: ...Tree 2 Right click on that protocol flag number 3 From the menu that is displayed select Associate Assay The Associate Assay window is displayed 4 Select the Assay that you would like to associate to...

Page 28: ...ing either an RTF or Delimited Text file for the assay and would like the data to be saved in a directory other than the default Check the Use Default Output Data Path box if you would like the data t...

Page 29: ...cassette Figure 2 13 Cassette 3 Place all of the background reference and sample vials into the appropriate cassette positions 4 Place the cassette s in the sample changer deck so that the protocol nu...

Page 30: ...1 In the Protocols Tree of the main window select the assay you would like to disassociate from a protocol number Figure 2 14 Protocol Tree 2 Right click on that assay Note You can also disassociate a...

Page 31: ...HOWTO PerkinElmer Life and Analytical Sciences 21 How to Edit an Assay 1 Select the File Open Assay menu option The Open Assay window is displayed Figure 2 15 Open Assay Window...

Page 32: ...2 16 Assay Definition Window 3 In each of the seven tabs of the Assay Definition window change the assay parameters as needed 4 If you would like to save the assay with the same name click the OK but...

Page 33: ...rint window is displayed 3 In the print window select the appropriate print parameters and click the OK button A list of the Assay Parameters Count Conditions Count Corrections and Reports defined for...

Page 34: ...to a Sample Nuclide Note Refer to page 26 on how to set up a quench set standards assay This requires the single dual color DPM option on the 2800TR 1 Select the Libraries Sample Nuclides Main menu i...

Page 35: ...ne nuclide in one counting region the Quench Set Medium if you are counting a second radionuclide in a separate region and the Quench Set High if you are counting a third radionuclide in a third regio...

Page 36: ...ssary 3 Save the assay in the Assays folder of the Packard TriCarb directory 4 Associate link the assay parameters with a protocol and attach the corresponding protocol flag to the first cassette to b...

Page 37: ...use the discriminator setting from the Standard Set 4 Create a new assay by choosing Alpha Beta as the assay type On the Count Conditions tab select the desired alpha beta radionuclide name from the...

Page 38: ...he drop down menu OR Open an existing assay and edit or review if necessary 6 Click the Name button on the Count Conditions Tab to choose from the list of Alpha Beta Standard names in the Alpha Beta S...

Page 39: ...the main window click on the Replay tab 2 Select the results file that you would like to analyze A listing of the results files is displayed in the Replay Tree using file names with the following synt...

Page 40: ...nditions tab define the parameters as needed for reanalysis of sample data 6 In the Report Definition tab define any printed or electronic reports you would like to generate for the reanalysis of samp...

Page 41: ...ure 2 21 Radionuclide Decay Window 2 From the drop down list in the Nuclide field select the nuclide of interest if it appears in the list select manual if you would like to manually enter the nuclide...

Page 42: ...Start Programs Administrative Tools Common and the click on User Manager 4 Click on the Policies menu and select User Rights The User Rights Policy window will display In the Right field use the drop...

Page 43: ...communications port a parallel printer port and a graphics adapter for the color monitor All RS 232 communication occurs through communications port one Communications port two is used for proprietary...

Page 44: ...CHAPTER 3 34 PerkinElmer Life and Analytical Sciences...

Page 45: ...ns in the Protocols Tree and data analyses in the Output Report and SpectraView windows Any assay data that is collected may be reanalyzed without recounting samples using the system s optional Replay...

Page 46: ...PerkinElmer Life and Analytical Sciences Main Window The main software window is comprised of several functional elements which provides you with a means of accessing all instrument features Figure 4...

Page 47: ...The Output Window is a multipurpose window which displays various data items and assay parameters It is a copy of the printer report if one is selected The output is defined in the Report Output wind...

Page 48: ...vides you with information about the status of a sample count and the region settings used in the counting procedure A number of display options are available for the spectrum and are defined in this...

Page 49: ...rding the status of a current protocol and displays instrument messages Figure 4 4 Instrument Status Bar The buttons in the Instrument Status Bar of the main software window allow you to start stop an...

Page 50: ...ers and the assay names that you have associated to flag numbers Existing reports are also shown in this view During protocol execution this window uses different symbols to provide a visual indicatio...

Page 51: ...eted but that the protocol has remaining count cycles This yellow flag will also appear if you interrupt the protocol for a Priostat operation Figure 4 8 The Yellow Flag A checkered flag is displayed...

Page 52: ...11 The Yellow Prohibitory Symbol The page symbol indicates that a report has been defined for this assay The report name appears next to this symbol in the Protocols Tree Figure 4 12 The Page Symbol...

Page 53: ...ontaining previously collected data which can be reanalyzed using different data reduction conditions without recounting the samples Figure 4 14 Replay Tree Note Clicking twice on the Results folder w...

Page 54: ...n item to display a list of options within that menu Note Some menu items represent optional features If your instrument is not equipped with the optional features the corresponding menu items will be...

Page 55: ...STEM SOFTWARE PerkinElmer Life and Analytical Sciences 45 File Menu Display the File menu by selecting File from the Menu Bar Figure 4 17 File Menu New Assay This menu item allows you to define a new...

Page 56: ...menu item allows you to Associate link an assay that you have defined to a protocol number Disassociate Assay You can disassociate an assay from a protocol flag number by selecting the assay you would...

Page 57: ...ose to save data to the default directory by checking the Use Default Output Data Path box Figure 4 19 Data Paths Window Note This information is stored on a per protocol basis as indicated in the Win...

Page 58: ...is also available via the red button on the Instrument Status Bar Start Resume Counting This menu item loads the sample currently at the detector position and begins counting This command is also ava...

Page 59: ...unting of the Priostat priority samples and resumes counting of the interrupted samples Sample Priostat not available on the 2800 optional on the 2900 TriCarb This is a special interrupt mode which al...

Page 60: ...al Preview The menu item allows you to view the sample spectrum and approximate the activity for a sample using Normal count mode Low Level Preview This menu item allows you to view the sample spectru...

Page 61: ...t quench curves for calculating DPM Disintegrations Per Minute in DPM Assays Quench Standards are counted once and the entire spectrum for each quench standard is stored independent of assay informati...

Page 62: ...e and save nuclide names and counting region limits for radionuclides You can also select a quench set for each sample nuclide using the Quench Set buttons in this window In Replay use the Sample Nucl...

Page 63: ...s is used to construct quench curves for determining sample DPM Disintegrations Per Minute Note Name max keV and DPM are the only entries required when entering a new quench set The remaining paramete...

Page 64: ...llows you to display the Alpha Beta Nuclides Library window The Alpha Beta Nuclides Library is a repository of information regarding previously defined Alpha and Beta emitting nuclides counted in Alph...

Page 65: ...isplay the Alpha Beta Standards Library window The Alpha Beta Standards Library is a repository of information regarding previously defined standards used in Alpha Beta Assays For additional informati...

Page 66: ...ions Per Minute DPM of radionuclides Figure 4 28 Radionuclide Decay Calculator Options Select this menu option to display the Options window The Options window allows you to select a format for the Sp...

Page 67: ...isplay certain error messages that are associated with an assay Figure 4 30 Show Protocol Errors Spectral Mapping The Spectral Mapping menu item allows you to view a three dimensional spectral map for...

Page 68: ...FS DPM sample counting procedure Figure 4 32 Spectrum Unfolding IPA Menu Display the IPA Instrument Performance Assessment menu by selecting IPA from the menubar Figure 4 33 IPA Menu The IPA menu can...

Page 69: ...stics Menu Display the Diagnostics menu by selecting Diagnostics from the menubar The TSE Diagnostics item in the Diagnostics menu is for the use of a PerkinElmer Service Engineer to view the system s...

Page 70: ...menu allows you to define a format for the window display on the monitor and restore the SpectraView window Figure 4 35 Window Menu Cascade This menu item displays the open windows in the following f...

Page 71: ...he spectrum for the current sample Help Menu Display the Help menu by selecting Help from the menu bar Figure 4 38 The Help Menu Help Topics This menu item will launch the On Line Help documentation w...

Page 72: ...ee dimensional view The X axis of the map represents the energy in keV the Y axis represents the counts and the Z axis represents the quench indicating parameter tSIE The spectral map can be used for...

Page 73: ...Spectrum Unfolding window can display three dimensionally the composite nuclide spectrum as individual separate spectra for each nuclide The X axes of the spectral displays represent the energy in keV...

Page 74: ...Figure 4 41 The SpectraView Window The SpectraView window is typically used for the following Monitoring sample counting Detecting spectral distortions or compressions resulting from sample quench Obs...

Page 75: ...the Report Definition tab of the Assay Definition window first define a named report with the information that you need On the Report Output tab select the desired named report and mark the Output to...

Page 76: ...a file format Other predefined sets of data can also be selected for output using the Special Files tab of the Assay Definition interface This other data includes Composite Spectra File Individual Sp...

Page 77: ...terference such as quench Performing a CPM Assay The following tasks are required when performing a CPM Assay Calibrate the instrument if necessary Create a new assay choosing CPM as the assay type an...

Page 78: ...hree nuclides in a sample using one two and three defined counting regions FS DPM Full Spectrum DPM Assays allow you to count two nuclides in a sample using the full spectrum of the sample no defined...

Page 79: ...ull Spectrum DPM method is used for regionless counting of dual label samples It uses the Quench Indicating Parameters Spectral Index of the Sample SIS and transformed Spectral Index of the External s...

Page 80: ...ith a protocol and attach the corresponding protocol flag to the first cassette to be counted Load the cassette s with vials and load the instrument with cassettes Start the instrument Save the assay...

Page 81: ...etween 20 and 40 then the DPM is reported and the sample is designated as indeterminate I on the printout If the SIS value is less than or equal to 20 and the sample is determined to contain Tritium t...

Page 82: ...rimination between the nuclides difficult To separate alpha and beta nuclide energies you must take advantage of their differing pulse decay times The TriCarb instruments equipped with the Alpha Beta...

Page 83: ...nator setting from the Standard Set If you do not choose a Standard Set manually enter a discriminator setting directly in the Alpha Beta Nuclide Library Open an existing assay or create a new assay b...

Page 84: ...cally included by default in alpha beta assays Beta activity can be reported in 2 regions CPMA and CPMB CPMA is always included by default in the report 4 High sensitivity HSCM or Low Level Count mode...

Page 85: ...less than 50 000 CPM choose Manual as the Discriminator Type by clicking on it The remaining fields are for information only They are either default values or values computed by the instrument Choose...

Page 86: ...you would like to count Performing a Quench Standards Assay The following tasks are required when performing a Quench Standards Assay Calibrate the instrument if necessary Create a new assay choosing...

Page 87: ...important to reduce the instrument background to its lowest possible level Lowering the high voltage supplied to the photomultiplier tube typically decreases background and increases sensitivity in SP...

Page 88: ...assay Define a sample nuclide in the sample nuclides library if one does not already exist Link the nuclide to the assay Link standards to the assay if necessary Specify the appropriate count conditi...

Page 89: ...this window Figure 5 1 Assay Definition Assay Parameters Tab Save As Click this button to save any changes made in the Assay Definition tabs using a different filename or file location on the disk Pa...

Page 90: ...represents the date the assay was created Date Modified This field represents the date the assay was last modified Lock Assay Mark this box if you would like to restrict editing functions for this as...

Page 91: ...Life and Analytical Sciences 81 Count Conditions The Count Conditions tab in the Assay Definition window allows you to define specific counting parameters for an assay Figure 5 2 Assay Definition Cou...

Page 92: ...ist of available count modes will be different Normal mode is the default and works well for most samples High sensitivity provides higher sensitivity as a result of the strict criteria used to exclud...

Page 93: ...dual and triple label experiments with variable quench samples where optimal region settings are desired SIS Spectral Index of the Sample SIS assigns a numeric value to the quench associated with a s...

Page 94: ...ffers from Assay Count Cycles where samples are unloaded from the detector and make a complete cycle around the sample changer deck prior to recounting Calculate Percent Reference Activate the percent...

Page 95: ...round values established during the calibration and Instrument Performance Assessment IPA procedures from the entire spectrum of the samples The background spectra are stored during these procedures a...

Page 96: ...nter the level of statistical accuracy for each region as a percent value that you would like to achieve before counting terminates Counting terminates when the sigma value for each region is reached...

Page 97: ...and Analytical Sciences 87 Count Corrections The Count Corrections tab in the Assay Definition window allows you to define specific count correction parameters for an assay Figure 5 4 Assay Definitio...

Page 98: ...herefore it is typically not used in this assay type It is especially important in SPC Assays to employ the above mentioned techniques to minimize the likelihood of generating static Luminescence Corr...

Page 99: ...old produce slower decaying pulses which may require longer delay times When using these scintillators it is advisable to lengthen the delay time to retain high counting efficiencies This is especiall...

Page 100: ...me The Report Name field allows you to assign a descriptive name to a report Add Report Click this button to define a new report for the assay You can choose to use any of the named report formats in...

Page 101: ...nding custom or data field to appear on the report Note Any changes to the format or equation of custom fields will be saved when the Assay Definition window is closed whether or not the Use field is...

Page 102: ...formula using mathematical operands on other field values and or constants Only one operation per Custom Field may be defined For complex operations you may find the need to create several Custom Fiel...

Page 103: ...ppropriate is NOT counted as a digit for this purpose The number of digits to the right of the decimal point can also be defined in the same manner Click OK to save any changes For fields that may res...

Page 104: ...field in the drop down list marked Field One Operand Click on the One Operand radio button to select from a list of unary operators operators that work on a single operand and a single field from the...

Page 105: ...s are available You can type in a numeric constant either an integer or a floating point number into the field or select from the entries in the drop down list Operator This drop down allows you to se...

Page 106: ...aView Block Data IPA Data Block Data Assay Definition Quench Curve Block Data I D Instrument Data Block Data Block Data items can also be used to send files to an external device via the RS 232 commun...

Page 107: ...ew Block Data This Block Data item allows you to print a graphical display of the spectrum for each sample Report Field Order Displays the order and format in which the report fields will be output Fi...

Page 108: ...rt Figure 5 10 Assay Definition Report Output Tab Report Name This box allows you to select a named report for output Output To Printer Select this to send the report to the printer Data from an assay...

Page 109: ...written from one run to the next Delimited Text Generates data as delimited ASCII text If you enter a numerical extension this selection is identical to the Auto Incrementing selection described above...

Page 110: ...tion included in the data file output RS 232 Check this to allow QuantaSmart to transmit lines of sample data via the RS 232 port Choose either Output per Sample and or Average Output per Average allo...

Page 111: ...am to further process your data If information is entered in any of the fields but the Run Application checkbox is unchecked the information in the fields is retained but they are grayed out Applicati...

Page 112: ...tch the application program will run once at the end of the protocol Cycle the application program will run at the end of each cycle Sample the application program will run after each sample Special F...

Page 113: ...in it The data file format is similar to that created for individual spectrum files as shown in with count parameters for a single sample in a header and the channel data listed serially after it The...

Page 114: ...data files are created with a filename that corresponds to the following structure 3HCPM_20040511_1024_S0001_R01 Spectrum where 3HCPM protocol name 20040511 the year month and day 1024 the hour and m...

Page 115: ...s QuantaSmart to generate a 2000CA dat file an ASCII comma delimited text file that contains information about a protocol Worklist The Worklist tab in the Assay Definition window allows you to designa...

Page 116: ...Row Click this button to delete selected rows from the Sample Order table Delete All Click this button to delete all rows from the Sample Order table Sample Library Click this button to display the Sa...

Page 117: ...rklist The Export Worklist window is displayed Enter a descriptive name for the Worklist and click Save The exported worklist file is saved in the required format so that you can later use the Import...

Page 118: ...CHAPTER 5 108 PerkinElmer Life and Analytical Sciences...

Page 119: ...ored independent of assay information This allows you to select and use the same quench set in any number of assays and construct a quench curve for each sample at the time the sample is counted The S...

Page 120: ...Note The fields that are enabled in the Sample Nuclides Library will be different when accessed from different locations within the software The list of nuclides that is displayed is dependent on the...

Page 121: ...limit for region C measured in keV C UL Upper Limit Region C This field represents the upper counting limit for region C measured in keV C Half Life Half Life of Nuclide C This field represents the h...

Page 122: ...rom a sample nuclide right click on the corresponding quench set button in the Sample Nuclides table and select Clear Quench Set Comment Click this button if you would like to enter descriptive commen...

Page 123: ...d is stored in the Quench Standards Library This allows you to select and use the same quench set in any number of assays In Replay the Quench Standards window allows you to select a quench standards...

Page 124: ...his is the name given to the alpha beta nuclide To type in a new name click on the ADD button This will create an additional row to define a new nuclide Type in the desired name to be used in an alpha...

Page 125: ...or the Beta nuclide in Region B in keV Note This value cannot exceed 2000 Beta B UL This is the upper limit of the counting region for the Beta nuclide in Region B in keV Note This value cannot exceed...

Page 126: ...u can enter the discriminator value manually Count Mode If an Alpha Beta Standard Set is referenced by the nuclide then Normal Count Mode is displayed If no Alpha Beta Standard set is linked then this...

Page 127: ...he upper limit of the counting region for the Alpha nuclide in keV Note This field is for information only in this library Discriminator Type Select either Automatic or Manual from the drop down list...

Page 128: ...ed to count the Alpha and Beta Standards Note This field is for information only Coincidence Time This field represents the Coincidence Time of 18 s used when counting the Alpha Beta Standards Assay N...

Page 129: ...ces 119 Delete Click this button to delete a selected entry from the table Comment Click this button if you want to enter descriptive comments The Comment window will appear AB Nuclides Click this but...

Page 130: ...able and are also plotted graphically The calculated optimum discriminator setting appears as the Computed value in this window If you want to further minimize misclassified events for one of the nucl...

Page 131: ...lied to each of the two Photomultiplier Tubes PMTs is adjusted until the tubes have been synchronized in their response to a Carbon 14 standard Then the high voltage to both PMTs is adjusted simultane...

Page 132: ...perform the SNC IPA procedure If 23 hours have not elapsed since the previous calibration and normalization the SNC IPA cassette is bypassed and this procedure is not performed Ideally the instrument...

Page 133: ...tional feature on the 2800 and 2900 series instruments IPA Definition Before performing any Instrument Performance Assessment IPA you must define the parameters that the instrument will use in the ass...

Page 134: ...Typically a 60 minute count time is used to assess background Tritium Figure of Merit Threshold This is the lower limit for the calculated Tritium Figure of Merit value If the figure of merit value f...

Page 135: ...le of Datapoints to Establish Baselines Enter the number of IPA runs you would like to use to generate baselines for Tritium and Carbon 14 background and efficiency The range for this value is between...

Page 136: ...xceeded a message is displayed in the main window Reset Baselines Click this button to delete the current baselines The instrument will establish new baselines from the number of datapoints indicated...

Page 137: ...to view edit and print data for individual IPA parameters in chart or tabular form When the data is displayed as a table points can be included or excluded from the data set The axes on the chart rep...

Page 138: ...le sample in the detector 20 consecutive times with a count time of 30 seconds for each repeat measurement The normal range for this value is 7 63 to 36 19 with 95 confidence A properly performing ins...

Page 139: ...s parameter are recalculated as new IPA data are generated The limit for this parameter is baseline minus 3 Tritium Efficiency Baseline This chart displays the Tritium efficiency data based on a fixed...

Page 140: ...g messages Any performance parameter that fails to meet specification will generate a warning message at the top of the report Any IPA or calibration operation that fails to complete satisfactorily wi...

Page 141: ...t the SNC IPA protocol will only run if 23 hours have elapsed since the last time that the protocol was run Note The cassette must be loaded in the following order Figure 7 4 Standard SNC IPA Cassette...

Page 142: ...a sample Load the calibration cassette after the last cassette belonging to the current protocol When the current protocol is completed the calibration cassette will automatically move into the counti...

Page 143: ...t run Note The cassette must be loaded in the following order Figure 7 5 Super Low Level SNC IPA Cassette Loading 2 Load the purged unquenched Carbon 14 standard supplied into the first position of th...

Page 144: ...t counting Load the calibration cassette on the right hand side of the sample changer deck such that you can read the protocol flag Press the Start button to begin counting If the instrument is curren...

Page 145: ...t from the Run menu The Associate assay for Priostat window is displayed 3 Select the assay you would like to use to count the samples Note The interrupted protocol number cannot be selected to count...

Page 146: ...egion allows you to optimize sample counting region settings for a sample and determine the equivalent unquenched region settings for variable quench samples Low Level Optimize allows you to optimize...

Page 147: ...f the nuclide is long in relation to the total count time the radioactive sample is luminescent You can view either the sample spectrum or the results plotted as a histogram The histogram displays the...

Page 148: ...enu SPC Decay The SPC Single Photon Counting feature allows you to measure the rate of decay of luminescence for a nonradioactive sample In SPC counting a single photomultiplier tube is used to collec...

Page 149: ...the Run menu The Sample Priostat window is displayed 3 Select Count from the Run menu Sample counting commences immediately after the instrument identifies the Priostat flag 4 To stop the counting of...

Page 150: ...alculated SIS and tSIE Quench Indicating Parameters As a result of this relationship the Identify Nuclide feature allows you to identify any of the following nuclides in a single label sample Carbon 1...

Page 151: ...is displayed 3 Select Count from the Run menu Sample counting will commence immediately after the instrument identifies the Priostat flag 4 Stop the counting of a sample by selecting Stop from the Ru...

Page 152: ...If you would like the instrument to automatically optimize the sample counting regions with respect to the chemical environment of the sample you can perform the Reverse Region compensation procedure...

Page 153: ...or a background sample is ten minutes 5 Stop the counting of a sample by selecting the Run Stop menu item 6 Select Next Sample from the Run menu if you would like to count additional samples 7 Stop th...

Page 154: ...ting region settings such that they take into account your sample s chemical environment Reverse Region compensation should be used Reverse Region determines region settings using a factor based on th...

Page 155: ...vel and upper level region settings by observing the sample spectrum 5 Stop the Sample Priostat count via the Run Stop menu option Select Next Sample from the Run menu if you would like to count addit...

Page 156: ...s and thus increase the overall Figure of Merit for a counting procedure using the Low Level count mode If you would like to also optimize the sample counting regions with respect to the chemical envi...

Page 157: ...statistical accuracy 2S of 2 00 equivalent to 10 000 gross counts before terminating the counting procedure The recommended minimum count time for a background sample is ten minutes 5 Stop the countin...

Page 158: ...lpha Beta Preview function is a Sample Priostat option available via the Run menu This feature allows you to observe the sample spectrum and approximate the count rate of a sample containing an Alpha...

Page 159: ...2 Select Alpha Beta Preview from the Run menu The Sample Priostat window is displayed 3 Select Count from the Run menu Sample counting will commence immediately after the instrument identifies the Pr...

Page 160: ...orm the following tasks 1 Load your sample s into a cassette Attach the Priostat flag in the reset position to the cassette 2 Select Normal Preview from the Run menu The Sample Priostat window is disp...

Page 161: ...d to perform the following tasks 1 Load your sample s into a cassette Attach the Priostat flag in the reset position to the cassette 2 Select Low Level Preview from the Run menu The Sample Priostat wi...

Page 162: ...main window are the same as those listed in the Protocols tab Any modifications made to protocols during Replay do not affect the original acquisition parameters To access this function select the Rep...

Page 163: ...ies the AEC automatically monitors and adjusts the counting region to exclude unwanted background This setting is typically used for dual and triple label experiments with variable quench samples wher...

Page 164: ...ground values subtracted from each sample within each region of the assay Apply Half life Correction Mark this box if you would like to activate the Half life Correction feature When this feature is a...

Page 165: ...ndow The Output Window displays various assay parameters and data items You may customize the information that is displayed in this window by defining the Output Window Printer report This report is d...

Page 166: ...he amount of electronic background discrimination For High Sensitivity and Low Level Count Modes the criteria employed for discriminating true Beta counts from background counts becomes more stringent...

Page 167: ...it is necessary to secure a source of water for the background samples which is dead or free of radioactivity Samples must be free of extraneous sources of radionuclides such as that which may be fou...

Page 168: ...nce for a given location must be determined at the site of the installation Caution When using the High Sensitivity or Low Level Count Modes you MUST NOT use quench standards which have been purged fr...

Page 169: ...discrimination between the nuclides impossible on the basis of energy discrimination However alpha induced pulses in a scintillation cocktail have a longer duration than beta induced pulses Pulse shap...

Page 170: ...e associated for use with an alpha beta radionuclide in the Alpha Beta Nuclide Library By selecting the appropriate alpha beta radionuclide in an assay the PDD from the Alpha Beta Nuclide Library is u...

Page 171: ...that includes the desired data fields Specify that a data file of the desired format be created Select any appropriate special files that may be required by your application Prot dat and 2000ca dat ar...

Page 172: ...te This option does not apply to Tandem Processing in Replay Sample Select this button if you would like to run the application program after the data for each sample is collected Note Each time the a...

Page 173: ...e per week or whenever a system malfunction occurs 1 Check that the cassettes and sample changer surfaces are free of dirt 2 Cassette movement should be smooth with no excessive vibration 3 Check for...

Page 174: ...ch contains different types of program or data files Note Because the system requires that certain program files are located in specific folders the files in the sub folders should not be moved Assays...

Page 175: ...cassette is properly labeled with a protocol flag No Display 1 Confirm that the computer and video monitor power switches are in the ON position 2 Reboot the computer by simultaneously pressing the c...

Page 176: ...on Sample Changer Idle 1 Confirm that the instrument power switch is in the ON position 2 Determine if all sample counting cycles have completed 3 Check for an error message in the Status Bar of the m...

Page 177: ...er line for the instrument or call PerkinElmer Technical Service High Chi Square Problem with high voltage power supply or the photomultiplier tubes Call PerkinElmer Technical Service Note A properly...

Page 178: ...he time This is a result of the statistical nature of the Chi Square test Low Tritium Efficiency 1 Verify that the instrument has been calibrated If not perform an SNC calibration 2 Confirm that the c...

Page 179: ...rate within the first 30 seconds of the sample count message column If this character appears in the Messages column of a report the sample was determined to be heterogeneous Typically this is due to...

Page 180: ...ars in the Message column of a report this data line represents data about a Background Subtraction vial BACK PIN JAM FWD A Sample changer error has occurred No action is required unless a message is...

Page 181: ...l flag Cannot associate another assay to this flag while the current assay associated with this flag is active Multiple assays cannot be associated to the same protocol flag number Allow the instrumen...

Page 182: ...type cannot be used in Replay Only CPM Single Dual Triple and FS DPM assays may be used in Replay Can t drop vial A Sample changer error has occurred No action is required unless a message is printed...

Page 183: ...s the extrapolated quench value for the sample may not be accurate Elevator down A Sample changer error has occurred No action is required unless a message is printed If this message appears on a prin...

Page 184: ...ermine the DPM value for this sample When a sample is designated as indeterminate the DPM reported may be valid if the sample is not heavily quenched Check the tSIE value of the sample to determine th...

Page 185: ...ired unless a message is printed If this message appears on a printout call PerkinElmer Technical Service Missing Vial This message indicates that a vial is missing from the cassette No Quench sets sp...

Page 186: ...left photomultiplier tube detected inappropriate beta energy Check that the unquenched Carbon 14 source is in sample position 1 in the cassette Call PerkinElmer Technical Service Normalization error r...

Page 187: ...is displayed intermittently shut down Windows and then turn the instrument power off and back on If this message is not displayed wait and periodically retry the sample changer commands Priostat searc...

Page 188: ...sets One or more quench sets associated with a multiple label sample nuclide has been deselected from the assay or deleted from the library Re select the appropriate quench sets for the sample nuclide...

Page 189: ...ropriate cassette Warning C 14 Normalization DPM not defined The C 14 Normalization DPM has not been defined In the IPA Definition window enter a DPM value for the C 14 Standard Warning H 3 Normalizat...

Page 190: ...chnical Service Warning questionable H 3 chi square value please view historic data The Chi Square result is outside the range of 7 63 to 36 19 Verify the integrity of the Carbon 14 spectrum via the S...

Page 191: ...to the assay that is currently counting Allow the current assay to finish counting prior to making changes You cannot exit the application because the instrument is counting a protocol The QuantaSmart...

Page 192: ...CHAPTER 9 182 PerkinElmer Life and Analytical Sciences...

Page 193: ...any other calculations such as half life correction are performed The correction of the sample count for background is calculated as Background Threshold The background threshold value is expressed a...

Page 194: ...ns Per Minute Disintegrations Per Minute DPM is an expression of radionuclide activity DPM are calculated as follows Efficiency Counting efficiency is a measurement of the instrument s ability to quan...

Page 195: ...t Low Count Reject LCR is a means of rejecting samples if the sample nuclide activity doesn t meet a minimum specified criterion Samples are rejected and flagged if they do not reach the specified cou...

Page 196: ...tivity 2 Sigma 2s This value represents the percent of uncertainty in a gross count value with 95 confidence limits It is calculated as The following table indicates the total accumulated counts neces...

Page 197: ...ionuclide Decay Calculator Reference Ref This value represents the percent of a reference standard and is calculated using the CPM of the unknown sample and the CPM of the reference standard The refer...

Page 198: ...CHAPTER 10 188 PerkinElmer Life and Analytical Sciences...

Page 199: ...in metals glasses ceramics and concrete Becquerel Bq The Becquerel is the International System of Units basic unit of radioactivity One Becquerel One Disintegration Per Second One Curie 3 7x1010 Becqu...

Page 200: ...t of the chemical interaction of sample components The coincidence circuit used in the TriCarb instruments excludes most chemiluminescent events except when they occur at high rates Chi Square Test Th...

Page 201: ...lide or into a different energy state of the same nuclide This process results in a decrease over time of the number of original radioactive atoms in a sample It involves the emission from the nucleus...

Page 202: ...s two different radionuclides Efficiency Efficiency is the ratio of measured counts to the number of decay events which occur during a measurement interval It is usually expressed as a percentage Exte...

Page 203: ...clides or standards Luminescence This is a general term applied to the emission of light that cannot be attributed to the radioactivity of the sample It is typically due to a chemical reaction between...

Page 204: ...per keV of beta particle energy and usually a reduction in counting efficiency Quench Curve A quench curve is a mathematical graph which correlates the counting efficiency to a quench indicating para...

Page 205: ...t Library of frequently used sample names Scintillation Scintillation is a flash of light produced in a scintillator by an ionizing event The scintillation is the sum of all photons produced by the de...

Page 206: ...spurious pulses from the PMT Tandem Processing Tandem Processing is a process which allows a TriCarb instrument to pass data from a protocol to an application program TR LSC Time Resolved Liquid Scint...

Page 207: ...age of archaeological artifacts only the smallest sample may be spared for measurement without sacrificing the specimen In medicine and biomedical research improved sensitivity facilitates the use of...

Page 208: ...high energy cosmic radiation with the sample vial and photomultiplier PMT glass face Other non quenchable background interferences including PMT afterpulsing and natural radiation in the PMT and vial...

Page 209: ...THEORY PerkinElmer Life and Analytical Sciences 199 the background Figure 12 1 Liquid Scintillation Background Pulse Figure 12 2 Liquid Scintillation True Beta Pulse...

Page 210: ...necessary because Scintillation cocktails that contain solvents based on di isopropyl naphthalene DIN and 1 phenylxylylethane PXE have long scintillation decay constants i e produce light pulses of l...

Page 211: ...s maximum discrimination Progressing from NCM to LLCM yields increasing efficiency to background ratios E2 B The improvement is demonstrated in the table of 14 C data below PerkinElmer also offers the...

Page 212: ...guration Energy keV 14C Efficiency BKG CPM E2 B Ratio None Sample Only 20 0 113 0 64 99 3 62 1167 NCM Sample Only 20 5 101 5 64 03 3 28 1250 HSCM Sample Only 21 5 95 5 61 22 1 36 2756 LLCM Sample Only...

Page 213: ...urst Discrimination Modes TR LSC Count Mode Burst Discrimination Modes None Equivalent to conventional LSC systems NCM Minimum or Normal Count Mode HSCM Intermediate or Low Activity High Sensitivity C...

Page 214: ...any reagents used to convert the sample into benzene Overall the preparation of samples for extremely low level counting must be conducted with extreme care to provide contamination free samples for c...

Page 215: ...n counting low levels of 3H or 14C Certain plastic vials may also be used for large or small samples measuring 3 H in water It is recommended that all glass or plastic vials used for low level countin...

Page 216: ...dow should be NOT be marked Note Do NOT use High Sensitivity or Low Level Count Modes to count alpha samples in normal CPM or DPM counting assays These TR LSC discrimination modes are not appropriate...

Page 217: ...ght which makes up the pulse and determines the pulse decay time Beta nuclides typically produce a minimal slow component leading to a fast decay time Alpha nuclides produce a significant slow compone...

Page 218: ...e alpha emitter The composition of the standards must be as close as possible to the chemistry and volume of the subsequent samples to be counted When counting the pure beta emitter source the percent...

Reviews: