CALIBRATION
AND
NORMALIZATION
PerkinElmer Life and Analytical Sciences
121
Chapter 7
Calibration and Normalization
Before any samples are counted, you should calibrate and normalize the system
(Self Normalization and Calibration - SNC) and assess the performance of the
instrument (Instrument Performance Assessment - IPA) The SNC/IPA protocol
requires a special protocol flag and cassette loading that depends on your
instrument model.
Calibration/Normalization
During the calibration and normalization portion of the SNC, the voltage applied to
each of the two Photomultiplier Tubes (PMTs) is adjusted until the tubes have
been synchronized in their response to a Carbon-14 standard. Then, the high
voltage to both PMTs is adjusted simultaneously until the endpoint of the 14C
spectrum falls at the appropriate position in the Spectralyzer (on-board 4000
channel) multi-channel analyzer. This process is designed to ensure that the
instrument accurately quantifies the energy from all beta-particle emissions.
IPA
During Instrument Performance Assessment (IPA), the instrument measures
background, counting efficiency, sensitivity (Figure-of-Merit) and the
reproducibility of sample counting (Chi-Square test). Note: IPA is an optional
feature on the 2800 and 2900 series instruments.
Summary of Contents for QuantaSmart 1694267
Page 1: ...QuantaSmart For The TriCarb Liquid Scintillation Analyzer Reference Manual...
Page 9: ...v Chapter 12 Theory 197 Low Level Counting Theory 197 Alpha Beta Counting Theory 207...
Page 10: ...vi...
Page 44: ...CHAPTER 3 34 PerkinElmer Life and Analytical Sciences...
Page 118: ...CHAPTER 5 108 PerkinElmer Life and Analytical Sciences...
Page 192: ...CHAPTER 9 182 PerkinElmer Life and Analytical Sciences...
Page 198: ...CHAPTER 10 188 PerkinElmer Life and Analytical Sciences...