Prima 320 Advance
88
User Manual
Appendix
Guidance and manufacturer’s declaration - electromagnetic immunity
This device is intended for use in the electromagnetic environment specified below. The customer or user of the device should assure
that it is used in such an environment.
Required test
60601 test levels for equipment used
in a professional healthcare facility
environment only
Compliance level
Electrostatic discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2, 4, 8, 15 kV air
± 8 kV contact
± 2, 4, 8, 15 kV air
Radiated RF EM Fields
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
10 V/m
80 MHz to 2.7 GHz
Proximity fields from RF
wireless communications
IEC 61000-4-3
Frequency - Test level
385 MHz - 27 V/m
450 MHz - 28 V/m
710 MHz – 9 V/m
745 MHz – 9 V/m
780 MHz – 9 V/m
810 MHz - 28 V/m
870 MHz - 28 V/m
930 MHz - 28 V/m
1.72 GHz - 28 V/m
1.845 GHz - 28 V/m
1.97 GHz - 28 V/m
2.45 GHz - 28 V/m
5.24 GHz – 9 V/m
5.50 GHz – 9 V/m
5.875 GHz – 9 V/m
Frequency - Test level
385 MHz - 27 V/m
450 MHz - 28 V/m
710 MHz – 9 V/m
745 MHz – 9 V/m
780 MHz – 9 V/m
810 MHz - 28 V/m
870 MHz - 28 V/m
930 MHz - 28 V/m
1.72 GHz - 28 V/m
1.845 GHz - 28 V/m
1.97 GHz - 28 V/m
2.45 GHz - 28 V/m
5.24 GHz – 9 V/m
5.50 GHz – 9 V/m
5.875 GHz – 9 V/m
Electrical fast transient /
burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input / output lines
± 2 kV for power supply lines
Surge
IEC 61000-4-5
± 0.5, 1, 2 kV line(s) to earth,
± 0.5, 1 kV line(s) to line(s) for power supply
lines
± 2 kV line(s) to earth for input / output lines
± 0.5, 1, 2 kV line(s) to earth,
± 0.5, 1 kV line(s) to line(s) for power supply
lines
Conducted RF
IEC 61000-4-6
3 V - 150 kHz to 80 MHz
6 V - ISM bands between 150 kHz to 80 MHz
3 V - 150 kHz to 80 MHz
6 V - ISM bands between 150 kHz to 80 MHz
Voltage dips and interruptions
IEC 61000-4-11
0% UT (100% dip in UT) for 0.5 cycle
0% UT (100% dip in UT) for 1 cycle
70% UT (30% dip in UT) for 25/30 cycles
0% UT (100% dip in UT) for 250/300 cycles
0% UT (100% dip in UT) for 0.5 cycle
0% UT (100% dip in UT) for 1 cycle
70% UT (30% dip in UT) for 25/30 cycles
0% UT (100% dip in UT) for 250/300 cycles
Power frequency (50 Hz)
Magnetic field
IEC 61000-4-8
30 A/m
30 A/m