Prima 320 Advance
19
User Manual
Specification
4.3
Gas Supply Pressure
Pipeline
280-600 kPa (40.6-87.0 psig)
Cylinder (reduced pressure from
regulator)
400 kPa max
4.4
Flowmeter Tubes
Flow range
Air
0-15 L/min
Oxygen
0-10 L/min
Nitrous oxide
0-12 L/min
Accuracy
± 10% of full scale reading or ± 200 ml/min (whichever is greater)
4.5
Auxiliary Oxygen Supply
Supply pressure
250 kPa
Flow rate
100 L/min
Refer to Appendix 3 for details of a Penlon O2 therapy flowmeter kit
4.6
Auxiliary Common Gas Outlet (ACGO)
Connector
22 mm male conical connector incorporating a coaxial female 15 mm conical
connector.
4.7
Oxygen Flush
Flow rate
25-75 L/min
4.8
Fresh Gas Mixture
Anti-hypoxic fresh gas mixture
Minimum oxygen concentration : 25% +5%/-4% (of total O2 + N2O flow) - a minimum
of 21% oxygen
4.9 Environmental
Operating conditions
Temperature
10 - 40°C
Atmospheric pressure
70 - 106 kPa
Humidity
5 - 95% non-condensing
Transport and storage
Temperature
-20 to +55°C
Atmospheric pressure
50 - 106 kPa
Humidity
10 - 95% non-condensing
Cleaning and disinfection
Refer to section 7.4