Installation and Pre-Use Checks
6. Check that the cylinder valves are closed.
7. Make sure that the necessary emergency equipment is
available and in good condition.
8. Make sure that equipment for airway maintenance and
tracheal intubation is available and in good condition.
9. Make sure that the applicable anaesthetic and emergency
drugs are available.
10. Make sure that the castors are not damaged or loose and
the brakes are set and function correctly.
11. Make sure that the absorber unit is locked in position.
12. Make sure that the AC mains indicator and the battery
indicator come on when the power cord is connected to an
AC mains power source.
NOTE
If the AC mains indicator and the battery indicator are not
illuminated, the system does not have electrical power.
13. Make sure that the machine can be switched on or off
normally.
5.9.3 System check
1. Set the anaesthesia system switch (1) to the ON position.
2. Make sure that the yellow and red alarm lamps illuminate
3. Make sure that an audible beep sound is heard.
4. Make sure that the start-up screen is displayed.
5. Wait for the countdown on the Pre-use Test screen to
complete.
6. Make sure that the standby screen is shown after the Pre-
use Test screen.
7. Press the blue SYSTEM TEST button to start the pre-use
test
8. Select the Manual Leak Test button from the pre-use test
screen.
9. Follow the on-screen instructions before you press the
Start button.
10. When the manual and automatic tests are complete press
the FINISH button.
11. Press the QUIT button.
12. Set the APL valve to MIN.
13. Make sure that all gas flowmeter controls are turned fully
clockwise.
5.9.4 Power failure alarm test
1. Disconnect the machine from the mains power supply.
a) Make sure that the mains indicator is extinguished
b) Make sure that the battery indicator light illuminates.
1