Installation and Pre-Use Checks
5.2.3 Bag arm and bag installation
1. Fit the bag support arm (1) to the port (2) on the top face of
the absorber.
2. Turn the locking nut clockwise to secure.
3. Check that the bag arm can be rotated.
4. Fit the bag (3) to the bag arm outlet (4).
5.2.4 Oxygen sensor installation
CAUTION
1. Before installing the oxygen sensor (1), check that the
O-ring (2) is in good condition.
Fit a new sensor if the O-ring is missing or damaged.
2. Tighten the sensor manually, do not use a wrench or
other tool.
1. Align the oxygen sensor (1) with the port on the side of the
absorber.
2. Fit the sensor and turn clockwise to tighten.
3. Fit the cable (3) to the sensor.
WARNING
Make sure that the absorber canister position sensor (4)
is unobstructed by cables or tubing etc, at all times.
Do not route the oxygen sensor cable in front of the
canister sensor
4. Connect the cable (3) to the socket (5) on the side of the
anaesthetic machine.
5.2.5 Install the breathing hose and Y-piece connector
CAUTION
1. When installing the breathing hoses, hold the tube
connector at both ends of the tube to prevent damage to
the tube.
1. The two breathing hoses are connected to the expiratory
and inspiratory connectors.
2. Connect the breathing hoses to the Y-piece.
1
2
3
4
1
5
3
4
1
2