![Penlon Prima 320 Advance User Manual Download Page 27](http://html1.mh-extra.com/html/penlon/prima-320-advance/prima-320-advance_user-manual_1523898027.webp)
Prima 320 Advance
21
User Manual
Specification
4.12 Ventilator
Display
10.4“ TFT, with touchscreen
Waveforms
Pressure, Flow rate, Volume, P-V loop, F-V loop, Pleth (requires optional SpO2)
Support mode waveforms are displayed in colour
Drive gas
O2 or Air, user selectable
Automatic switching during operation, if there is a failure of the gas supply in use.
Inlet pressure
280-600 kPa
Max flow
≤
120 L/min
Ventilator modes
Volume control ventilation (VCV),
Pressure control ventilation (PCV)
Pressure-regulated volume controlled ventilation (PRVC)
Synchronised intermittent mandatory ventilation - Volume control (SIMV-V)
Synchronised intermittent mandatory ventilation - Pressure control (SIMV-P)
Synchronised intermittent mandatory ventilation - Pressure-regulated volume controlled
(SIMV-PRVC)
Spontaneous / Pressure support ventilation (SPONT/PSV),
Manual
Safety pressure
System pressure does not exceed 12.5 kPa
Parameters
Tidal volume
10 - 1500 mL
0 - 1600 ml measured in PCV
Increment: 10 ~ 100 mL: 5 mL
100 ~ 1500 mL: 10 mL
Respiratory rate
1
~
100 bpm; increment: 1 bpm
Tinsp (inspiratory time)
0.1 ~ 10.0 s; increment: 0.1 s
I:E ratio
4:1 to 1:10; increment: 0.5
Percentage of inspiratory pause 0 to 60%; increment: 5%
PEEP
OFF, 3 ~ 30 cmH2O; increment: 1 cmH2O.
Pressure support
0 ~ 70 cmH2O; increment: 1 cmH2O
Pressure control
5 ~ 70 cmH2O; increment: 1 cmH2O
Flow trigger
0.5 ~ 20 L/min; increment: 0.1 L/min
Pressure trigger
0 ~ 20 cmH2O; increment: 0.1 cmH2O
SIMV-P inspiration termination
level
5 ~ 80 %; increment: 5%
4.13 Monitored Parameters
Inspiratory tidal volume
0 ~ 2500 mL; Resolution: 1 mL
Expiratory tidal volume
0 ~ 2500 mL; Resolution: 1 mL
Minute ventilation (Mv)
0 ~ 60 L / min; Resolution: 0.1 L/min
Spontaneous minute ventilation
(Mvspont)
0 ~ 60 L/min; Resolution: 0.1 L/min
Respiratory rate
0 ~ 100 bpm; Resolution: 1 bpm
Spontaneous respiratory rate
0 ~ 100 bpm; Resolution: 1 bpm
I:E
30:1 to 1:150; resolution: 0.1
Peak airway pressure
0 ~ 100 cmH2O; Resolution: 1 cmH2O