![Penlon Prima 320 Advance User Manual Download Page 44](http://html1.mh-extra.com/html/penlon/prima-320-advance/prima-320-advance_user-manual_1523898044.webp)
Prima 320 Advance
38
User Manual
Installation and Pre-Use Checks
4. Check that a low pressure alarm is triggered.
5. Connect the Y-piece to the test lung.
6. Check that the low pressure alarm is cancelled.
5.12 Breathing System Test
1. Check that the non-return valves in the breathing system
work correctly:
a) The inspiratory non-return valve opens during
inspiration and closes at the start of expiration.
b) The expiratory non-return opens during expiration and
closes at the start of inspiration.
Bellows test
2. Set the system to standby mode.
3. Disconnect the Y-piece.
4. Set all gas flowmeter controls fully clockwise
5. Deflate the bellows.
6. Connect the Y-piece to the absorber test block.
7. Press and hold the O2 flush button to fill the bellows.
8. The airway pressure gauge must indicates less than 15
cmH2O.
9. Release the O2 flush button.
10. If the bellows deflates within one minute,.
a) Remove the bellows and reinstall, refer to Section
7.3.7.
b) Repeat the bellows test.
Breathing system leak test
11. Set the Bag/Vent switch to the bag position.
12. Turn the APL valve to 50 cmH2O
13. Press the O2 flush button until the airway pressure gauge
reads 25 cmH2O.
14. Release the O2 flush button.
15. Make sure that the airway pressure gauge reading does
not reduce.
CAUTION
If a leak is suspected, check the bellows, breathing tube,
absorber, and connectors for correct function and security.
Retest the system. Do not use the machine if the circuit
continues to leak. Contact a Penlon-trained engineer.
APL valve test
16. Turn the APL valve to 70 cmH2O position..
17. Connect the Y-piece to the test plug.
18. Push the O2 flush button until the airway pressure gauge
rises to approximately 35 cmH2O.