80
EXIT MENUS
O2 MONITOR & SPIROMETRY
FRESH GAS COMPENSATION: ON
SPECIAL MODES
WAVEFORM
ALARM SETTINGS
GAS MIXTURE: O2+AIR
USER SETTINGS
SERVICE MENU
O2 Monitor & Spirometry
ESCAPE FROM MENU
O2 MONITOR: on
CALIBRATION: 100%
HIGH ALARM SET: 105
LOW ALARM SET: 18
SPIROMETER: on
SPIRO CALIBRATION: 0 L/min
off / on
(Toggle option
21 / 100%
(Toggle option)
19 -105
(Integer)
18 - 99
(Integer)
off / on
(Toggle option)
0 L/min / 10 L/min
(Toggle option)
off/on
(Toggle option)
Fresh Gas Compensation
ON / OFF
Special Modes
ESCAPE FROM MENU
SUPPORT MODE: none
TRIGGER: 5 L/min
SIGH TO BREATH RATIO: 1:50
Support Mode pick list
none
pressure
simv
smmv
Sigh to Breath
Ratio pick list
1:50
2:50
3:50
4:50
Trigger pick list
0.2 L/min
0.3 L/min
0.4 L/min
0.5 L/min
1.0 L/min
1.5 L/min
2.0 L/min
2.5 L/min
3.0 L/min
4.0 L/min
Waveform
ESCAPE FROM MENU
SECOND WAVEFORM: off
Second waveform pick list
off
vol. vs time
vol. vs press.
Alarm settings
ALARM MENU
ESCAPE FROM MENU
ALARM MODE : default
HIGH TIDAL VOLUME: off
VM MIN: 0.3 L
VM MAX: 0.9 L
VT MIN: 300 mL
VT MAX: 900 mL
APNOEA ALARM LIMIT: 0.3 cmH2O
ALARM VOLUME: 50%
default / user
(Toggle option)
off / on
(Toggle option)
0.0 - 7.4
(Integer)
0.1 - 7.5
(Integer)
10 - 1600
(Integer)
20 - 2400
(Integer)
0.3 - 3.5
(Integer)
50 - 100%
(Integer)
Gas mixture: O2+Air
O2+AIR
O2+N2O
O2+Xe
O2+He
User Settings
ESCAPE FROM MENU
SELECT SETTINGS
SAVE SETTINGS
BACK LIGHT LEVEL: 50%
VOLUME TYPE: tidal
Service
See next page
Select settings
ESCAPE FROM MENU
USER1: CCT1
USER2: CCT2
USER3: CCT3
USER4: CCT4
USER5: CCT5
ADULT DEFAULT
PAEDIATRIC DEFAULT
Save settings
ESCAPE FROM MENU
USER1: CCT1 CONFIRM: CCT1
USER2: CCT2 CONFIRM: CCT2
USER3: CCT3 CONFIRM: CCT3
USER4: CCT4 CONFIRM: CCT4
USER5: CCT5 CONFIRM: CCT5
Backlight level
0 - 100% (integer)
Volume type
tidal/minute (toggle)
Menu Structure
Main Menu
Summary of Contents for AV-S