9.3
Connect the pressure tube to 'Patient Pressure' connection on the rear panel of the
control module to the Pressure Sensing port on the rear of the absorber.
9.4
Connect the cable from the 'Interface' connector on the rear panel of the control
module and to the Spirometer connection on the rear of the absorber and the rear of
the Prima SP anaesthesia.
9.5
Attach the 'Fresh Gas Supply' hose from the absorber to the CGO of the PrimaSP
9.6
Connect a patient breathing circuit to the CO2 absorber and attach the patient
connector to a test lung.
9.7
Connect the Gas Scavenging System to the "Exhaust' connector on the rear of the
vent. (If no scavenge system is attached to the rear of the ventilator, a continuous
bleed during the inspiratory phase will be audible)
10
Standby (every 6 months)
10.1
Power On the Anaesthesia machine. Ventilator should switch 'ON'
10.2
Power Off the Anaesthesia machine. Ventilator should switch 'OFF'
10.3
Press the Power button on the ventilator.
10.4
Default selection screen appears and automatically defaults to adult mode after
approximately eight seconds
10.5
Note: standby mode is highlighted white in the bottom right of screen.
10.6
Set incorrect rate, i.e. increase Vt, and/or Rate controls.
Check 'Incorrect Rate Or Ratio' displayed on screen and 3 alarm tones sound. Do not
confirm settings, wait for default values to return.
11
Oxygen Analyser Function and Calibration Tests (every 6 months)
11.1
Connect a calibrated O2 Analyser into inspiratory connector.
11.2
Check Vent O2 sensor is inserted into the absorber O2 sampling point.
11.3
Remove canister.
Attach a short hose to inspiratory limb.
Use O2 flush for 20 seconds the set a 5 l/min flow of oxygen and allow reading to
stabilise.
Ensure 100% indicated on test device and ventilator display is stable.
11.4
Calibrate Vent in 100% Oxygen.
11.5
From menu select 'Oxygen Monitor & Spirometry'-'Calibration'.
Adjust until 100% is indicated then press navigator wheel to confirm.
11.6
Exit menu and set fresh gas flow to minimum
11.7
Expose both sensors to air and check reading is 21% +/- 2
11.8
On ventilator, adjust high and low O2 alarms and check alarms trigger when values
SERVICE SCHEDULE