4.15
Oxygen Monitor
Measurement Range:
0-l00%
Resolution:
±1%
Accuracy and Linearity:
±2% of full scale (at constant temperature and pressure)
Response Time:
90% of final value in approx. 10 seconds (air to
100% O
2
)
Operating Temperature:
50°F to 100°F (10°C to 38°C)
Storage Temperature:
23°F to 122°F (-5°C to 50°C)
Relative Humidity Range:
5%-95% (non-condensing)
Battery Back-up:
As per ventilator
Sensor Type:
MOX-3 galvanic fuel cell
High Priority Alarm:
Flashing, 5 audio pulses with 6 seconds repeat
time.
Medium Priority Alarm:
Flashing, 3 audio pulses with 24 seconds repeat time
Low Priority Alarm:
Static with single beep sound
Alarm Mute:
30 seconds for high priority alarm
120 seconds for medium priority alarm
Low Alarm Set Range:
18%-99% (
±
1%)
High Alarm Set Range:
19%-105% (
±
1%)
Cable length:
2 m (6 ft), fully extended
Sensor
Type:
Galvanic fuel cell sensor (0-100%)
Life:
1500000 O
2
% hours
(One year minimum in typical applications)
Interference Gases and Vapours (in 30% Oxygen, 70% Nitrous Oxide)
Interference
Volume % Dry
Interference in O
2
%
Nitrous Oxide
80%
<1%
Carbon Dioxide
5%
<1%
Halothane
5%
<1%
Enflurane
5%
<1%
lsoflurane
5%
<1%
Sevoflurane
5%
<1%
SPECIFICATION
- O
2
Monitor
37