3.12.3
O
2
Monitor sub-menu
ON/OFF
Turn the navigator wheel to switch between
ON and OFF.
Press to confirm.
Scroll to EXIT MENUS and press the wheel
to exit.
NOTE
The oxygen monitor automatically switches ON
and defaults to the previous values for high and
low alarm settings when the ventilator is
switched on.
Fresh gas mixture compensation is disabled if the
O2 monitor is switched OFF.
CALIBRATION
Press the navigator wheel to initiate the
calibration procedure (see section 5.3.2 for
full procedure).
To exit the menu, scroll to EXIT MENUS
and press the wheel.
HIGH ALARM SET
LOW ALARM SET
Scroll to the required parameter and press
the navigator wheel to activate.
Rotate the navigator wheel again to change
the displayed value.
(see section 5.3.4 for full procedure).
High Alarm range:
19% to 105%
Low Alarm range
18% to 99%
The displayed figure will flash on and off.
Press to confirm.
Scroll to EXIT MENUS and press the wheel
to exit.
O
2
Monitor sub-menu
O
2
Monitor sub-menu - calibration
O
2
Monitor sub-menu - alarms
DESCRIPTION
32
O2 Monitor & Spiro
ESCAPE FROM MENU
>
O2 MONITOR: on
CALIBRATION: 100%
HIGH ALARM SET: 105
LOW ALARM SET: 18
SPIROMETER: on
SPIRO CALIBRATION: 0 L/min
O2 Monitor & Spiro
ESCAPE FROM MENU
O2 MONITOR: on
>
CALIBRATION: 100%
HIGH ALARM SET: 105
LOW ALARM SET: 18
SPIROMETER: on
SPIRO CALIBRATION: 0 L/min
O2 Monitor & Spiro
ESCAPE FROM MENU
O2 MONITOR: on
CALIBRATION: 100%
>
HIGH ALARM SET: 105
LOW ALARM SET: 18
SPIROMETER: on
SPIRO CALIBRATION: 0 L/min