
PRE-OPERATION PROCEDURES
46
B
X
C
5.1.11 Pressure Monitor Connections
WARNING
The High and Low Airway Pressure Alarms are
important for patient care.
The connection point must be properly
located in the expiratory limb of the breathing
system.
1.
PATIENT PRESSURE port (
A
) on the
rear panel of the control unit:
Use the appropriate Penlon tubing
assembly to connect to the expiratory
limb of the breathing system, close to
the circle system expiratory valve.
2.
Push-fit, self sealing connectors (B)
Push in the tube as far as possible
Do not use excessive force.
The connector end piece ‘X’ will also
move inwards.
Pull the tube carefully outwards.
The end piece ‘X’ will be pulled
outwards to the ‘locked’ position.
3.
Connect the tubing (with adaptor, Part
No 053049) to the push-fit, self-sealing
connector (
C
) at the rear of the
A200SP Absorber.
A
Summary of Contents for AV-S
Page 75: ...PARTS LIST 69 ...
Page 77: ...71 ...
Page 78: ...PARTS LIST 72 ...
Page 81: ...PARTS LIST 77 ...
Page 93: ......