On-screen Menus
79
APPENDIX 2
NOTE:
1.
All selection or changes in the menu are
followed by a "CONFIRM" message
prompt on the screen, and accompanied
by a "BEEP" (user volume set)
2.
The selected text or option will invert in
colour
3.
User settings menus only activate in
Standby mode.
4.
Clock menu, Upgrade menu, Diagnostic
menu only activate in Standby mode.
5.
Special Modes on-screen tab only
activates in Spontaneous mode
6.
Adult default settings
VT=600 mL
RATE=10 bpm
IE RATIO=1:2
Plimit=38 cmH2O
Ptarget=10 cmH2O
7.
Paediatric default settings
VT=150 ml
RATE=15 BPM
IE RATIO=1:2
Plimit=38 cmH2O
Ptarget=10 cmH2O