AV-S
Touchscreen control
1
4
3
5
2
Gas Mixture
O
2
+ air
.
.
IO
o
o
DESCRIPTION
5.
Navigator Wheel and Press Button
Turn the wheel to select a function or
parameter, or to alter the value of an
active parameter.
Press to confirm the setting.
3.5.1.2 Selecting Functions and
Parameters
The functions/parameters shown on the
screen can be activated as follows:
a) touch the screen at the appropriate tab
area.
b) rotate the navigator wheel and press it
when the indicator arrow is on the required
parameter tab
Note that parameters default to factory-set
values when the ventilator is switched on
and no further user selection is made.
3.5.1.3 User Adjustable Parameters
Variable parameters can be altered by
rotating the navigator wheel.
When the required value is displayed, press
the active tab
or
the wheel to confirm the
setting.
3.5
Control Panel
3.5.1
Touchscreen and Navigator
Wheel / Push Button
3.5.1.1 Control Panel
1.
On/Off control
Switch On:
Short internal test sequence
Switch Off:
5 second power down sequence with
audible tones
2.
Status indicators for electrical power
(mains/battery supply)
Yellow indicator
- illuminated whenever power is
applied to the unit and internal battery
is being charged
Green indicator
- illuminates when the unit is switched
on, .
3.
Menu switch
The menu function provides access to
user and service pages
4.
Alarm mute switch
30 second or 120 second Alarm
silence, depending on alarm status.
Note also that some alarms are not
mutable (see 3.11).