3.5.2
User Adjustable Parameters
Tidal Volume Range
20 - 1600 ml
Rate
4 - 100 bpm
I:E Ratio
1:0.3 to 1:8
PEEP
4 - 30 cmH2O
Can be set to OFF
Pressure Limit
Volume mode:
10 - 80 cmH2O
Pressure mode:
10 - 70 cmH2O
3.5.3
Operational Capability
Tidal Volume, Rate, and I:E ratio settings are all limited by a
maximum inspiratory flow of 75 L/min.
DESCRIPTION
17
0.1
0.2
0.4
0.3
0.6
0.5
1.6
1.5
1.4
1.3
1.2
1.1
1.0
0.9
0.8
0.7
0
10
20 30 40 50 60
70 80
1:6 1:5 1:4 1:3 1:2 1:1
1:0.3
Rate (bpm)
Tidal
Volume
(litres)
(Vt)
I:E Ratio
The ventilator is capable of operating at the volumes and rates below each I:E ratio curve.
Example
1.
Select required volume (Vt) (e.g. 0.8 L)
2.
Select rate (e.g. 10 bpm).
3.
Select I:E ratio of 1:2.
The point X on the graph lies beneath the 1:2 ratio curve, and is therefore within the
ventilator’s capability
.
X