Page 8
Technical Specification
Solaris XE
Environment
Solaris XE
Minimum Operating Ambient Temperature
5°C (41°F)
Maximum Operating Ambient Temperature
35°C (95°F)
Maximum Altitude
2000 m
Maximum Relative Humidity
80% Non-Condensing
Minimum Storage Temperature*
-20°C (-4°F)
Maximum Storage Temperature*
60°C (140°F)
Note: Please ensure that the generator is situated in a well ventilated environment.
Inlet Conditions
Inlet Air Pressure Min-Max
5.51 - 9.99 bar (80 - 145 psi)
(125psi min to achieve
max output of 35L/min)
Min Air Inlet Flow (For Max Outlet Flow)
up to 100 L/min
Minimum Air Inlet Quality
ISO 8573-1:2010 Class[1.4.1]
Generator Outlets
Maximum Output Pressure
116 psi / 7.99 bar (variable)
Maximum Outlet Flow Rate
up to 35 L/min
Purity
95-99.5% N
2
Particles
<0.01µm
Phthalates
NONE
Suspended Liquids
NONE
Gas Outlets
1
Start-Up Time For Purity
5 mins
Electrical Requirements
Voltage
100-230v
Frequency
50/60 Hz
Current
0.10A
Input Connection
C14 Plug
Power Cord (Supplied)
C13 Socket to Local Connection
Pollution Degree
2
Insulation Category
II
Fuses (UK Only)
13A BS1362
General
Dimensions cm (inches) WxDxH
34.3 x 65.5 x 15.6 (13.5 x 25.78 x 6.14)
Generator Weight Kg (lbs)
15.1 (33.2)
Shipping Weight Kg (lbs)
18 (39.6)
Max Weight Bearing Kg (lbs)
100 (220)