Vibrax VXR basic

20000005403

VXR_basic_122017

Summary of Contents for Vibrax VXR basic

Page 1: ...IKA Vibrax VXR basic 20000005403 VXR_basic_122017 ...

Page 2: ...ontens evenly distributed and well secured Because of improper loading and the position of the center of gravity dynamic forces may arise during the agitation process that cause the shaker to move about on the table In this case reduce the speed immediately Additional risk may be present for the user if flammable materials are used during the agitation process or if mechanical agitation energy is ...

Page 3: ...active force on the drive This force is comparable to an imbalance As a result there is an imbalance mass on the motor that compensates for a part of the imbalance forces Because of the different mass of each agitation material there is a residual imbalance This imbalance has the tendency to move the unit back and forth and to move it around on the adapter surface This is largely prevented by the ...

Page 4: ... circuit and the control light flashes To eliminate the error the weight of the material must be reduced allow the device to cool off The device must be turned off and backon again The IKA Vibrax VXR basic is maintenace free For cleaning disconnect the main plug It is subject only to the natural wear and tear of components and their statistical failure rate If the three silicon hoses that protect ...

Page 5: ...on stroke mm 4 Ø Shaking motion horizontal circular Perm duration of operation 100 Perm ambient temperature C 5 to 40 Perm relative humidity 80 Prozection class acc DIN 40 050 IP 21 Protection class II Overvoltage category II Contamination level 2 Operation at a terrestrial altitude m max 2000 above sea level Max load kg 2 including attachment Dimensions W x D x H mm 157 x 247 x 130 without attach...

Reviews: